OPERATION
Note:
Test the OTL system at least once per year for functionality.
S
ET THE
O
VER
T
EMPERATURE
L
IMIT
The incubator must be operating at your incubation application temperature and must be stable for at least
1 hour prior to setting the OTL. Allow the unit to operate for at least 8 hours prior to setting the OTL when
first placing it into operation.
If the OTL is sporadically activating, you may turn the dial very slightly to the right (clockwise).
If the OTL continues activating, check for ambient sources of heat or cold that may be adversely impacting
the unit temperature stability. Check if any powered accessories in the incubation chamber are generating
heat. If you find no sources of external or internal temperature fluctuations, contact
your distributor for assistance.
End of procedure
1. Set OTL control to its maximum setting,
if not already set to max.
2.
Turn the dial counterclockwise until the red Over Temperature Limit Light illuminates.
3.
Slowly turn the dial clockwise until the OTL Activated light turns off.
•
The Over Temperature Limit is now set
approximately 1˚C above the current
incubator air temperature.
4.
Leave the OTL dial set just above the activation point.
Optional:
Turn the dial slightly to the left.
•
This sets the OTL cutoff threshold nearer
to the current incubator air temperature.
Содержание SRI3
Страница 1: ...BOD INCUBATOR 110 120 Voltage 220 240 Voltage Installation Operation Manual SRI3 SRI3 2...
Страница 8: ...8 P a g e INTRODUCTION...
Страница 12: ...12 P a g e RECEIVING YOUR UNIT...
Страница 20: ...20 P a g e GRAPHIC SYMBOLS...
Страница 22: ...22 P a g e CONTROL PANEL OVERVIEW...
Страница 38: ...38 P a g e USER MAINTENANCE...