4
1.0 INTENDED USE
The SRPM is designed to monitor critical environments by providing
differential pressure indication. Typically this is between a monitored
room and a reference space such as a corridor or ante room. The unit
also provides monitoring, alarm and communications functions.
Installation must be indoors, Polution Degree 2, Installation Category II.
Typical Applications:
1. Hospitals – patient isolation and protection rooms, operating
suites, intensive care, and emergency rooms.
2. Pharmaceutical, semiconductor, precision manufacturing, and
clean rooms.
3. Laboratories – medical research, BSL (Bio safety labs), radiation,
vivarium, toxic metals and chemicals. Indoor use only.
1.1 SPECIFICATIONS
Service:
Air or nonconductive, nonexplosive gases.
Accuracy:
±0.5% F.S., ±0.25% FS optional
Operating Temperature Limits:
32 to 120°F (0 to 50°C).
Operating Humidity Limits:
5 to 95% Relative humidity
(non-condensing).
Altitude:
2000 meters (max).
Thermal Effects:
±0.03% F.S./ °F (± 0.05% F.S./ °C).
Overpressure:
±15” WC.
Supply Voltage:
Code V1 and V2 85-265 VAC, 50-60 Hz
Code A1 and A2 18-32 VAC, 50-60 Hz.
Power Consumption (Voltage output):
5 W.
Output:
Selectable 4-20 mA (2-wire), 0-5 VDC (3-wire), or 0-10 VDC (3-
wire).
Loop Resistance (4-20mA output):
0-510 OHMS
Ω.
Electrical Connection:
Removable terminal block.
Pressure Fittings:
Barbed fi ttings for 1/4” O.D. tubing.
Housing:
Fire retardant plastic (NEMA 1, IP20 rated for indoor
applications).
Mounting
: Mounts to customer supplied 4½ x 4½ plaster ring (mounts to
double gang electrical box).
Dimensions:
8½ H x 5.4½ W x 1.8½ D (20.3 H x 13.7 W x 4.1
Weight approx.):
1.5 lbs (680 g).
Agency Approval:
CE
, CAN/CSA-C22.2 NO. 61010-1-04
Communications:
BACnet MSTP ASC, optional, see setra.com for
detailed PICS statement.