User Interface DF-310E
81
When the composition of the gas (or the GSF factor) has been entered move the asterisk to
Update & Quit and press
. The GSF will be calculated and displayed.
If the GSF factor of the gas used to calibrate the system is already known, it can be entered
directly. To enter the GSF directly, move the asterisk to the
GSF
line and press
. Use the
and
keys and hit
to enter the desired value.
NOTE
The GSF for the gas used to calibrate the system may be different
from that used during analysis. If the GSF is changed to reflect the
composition of the calibrating gas, be sure to reset the GSF before
analyzing samples.