50 DF-310E
Connecting to External Devices
connections. See the wiring diagram in Figure 30 and refer to Table 10 and Table 11 for wire
sizes and lengths. Following are three remote sensor configurations and wiring diagrams.
Care must be taken when making up gas fittings on the sensor when mounted on a remote
bracket as shown in Figure 24 below. A backing wrench must always be used (in particular for
VCR connections) when connecting the gas sample line to the sensor. The inlet fitting, although
epoxied, is very delicate and the seal can easily be damaged if it is allowed to spin as the
connection is tightened. A 0.010 inch orifice is included with 0-50ppm sensors and must be
installed at the sensor inlet in order to control gas flow.
Contact Servomex for additional information on remote sensor installations.
7.5.1 Sensor on Remote Bracket with Optional Pump
CAUTION
Always use a backing wrench when connecting the gas sample line
to a remote sensor.
(WHT/BLK/RED)
SENSOR - (WHT/YEL)
SECONDARY ELECTRODE - (WHT/BLU)
SECONDARY ELE (WHT/RED)
FLOW SWITCH
FLOW SWITCH
UNUSED
PUMP +
GROUND
TEMP -
PUMP -
TEMP +
7.02 [178.28]
4.09 [103.89]
1.57 [39.88]
1.50 [38.10]
Ø.28 [Ø7.11]
4 PL
7.73 [196.29]
4.88 [123.83]
1
2
3
4
5
6
7
12
10
8
9
11
Figure 24: Remote Sensor with Optional Pump
Содержание DF-310E
Страница 1: ...Oxygen Analyzer DF 310E OPERATOR MANUAL...
Страница 10: ......
Страница 26: ...22 DF 310E Options Figure 7 Plumbing Configuration Options...
Страница 36: ......
Страница 62: ......
Страница 66: ...62 DF 310E User Interface 3 43 ppm...
Страница 96: ......
Страница 108: ......
Страница 118: ......
Страница 122: ......
Страница 124: ......
Страница 129: ......