24 DF-310E
Options
SAMPLE INLET
1/8"COMPRESSION
REGULATOR
232-GND
485-TX -
UNUSED
485-RX -
232-RX
232-TX
485-TX +
485-RX +
EXT-2 -
EXT+1 -
EXT-2 +
GND (H-)
SE -
SE+ (H+)
SNSR -
SNSR +
J7
UNUSED
TEMP -
TEMP +
UNUSED
UNUSED
LOOP -
J8
A4-NO
A2-NO
EXT-1 +
FLOW-B
UNUSED
FLOW-A
UNUSED
PUMP +
PUMP -
A OUT -
A OUT +
GND
A4-NC
GND
UNUSED
LOOP +
UNUSED
J5
UNUSED
UNUSED
24V RTN
UNUSED
+24V
A2-NC
J3
GND
J6
J4
A4-COM
UNUSED
A3-NC
A3-NO
A3-COM
A1-NC
UNUSED
A2-COM
A1-NO
A1-COM
J1
J2
Figure 9: Regulator Installation
5.8 Stainless Steel Outlet Tubing
Analyzers can be equipped with a 1/8-inch compression stainless steel outlet
tube. When this option is provided, the analyzer cannot be equipped with the
quick-disconnect fitting at the flowmeter outlet. Because of the rigid outlet
tube, the Sensor Assembly can only be removed after both inlet and outlet
bulkhead retainer nuts are removed. A 7/16-inch wrench is needed for the
inlet nut; and a ½-inch wrench is used on the outlet nut. When reinstalling the
sensor, make sure both bulkhead fitting hex sections are oriented to seat in the
retainer blocks on the inside rear of the enclosure.
5.9 Key Lock
An optional key lock can be installed in the door of the analyzer to prevent
access to the power switch and other internal components. The lock is
supplied with two keys.
If the analyzer is operating, the key lock does not prevent adjustments from
the front panel. Password Protection, described in the
User Interface
section
Содержание DF-310E
Страница 1: ...Oxygen Analyzer DF 310E OPERATOR MANUAL...
Страница 10: ......
Страница 26: ...22 DF 310E Options Figure 7 Plumbing Configuration Options...
Страница 36: ......
Страница 62: ......
Страница 66: ...62 DF 310E User Interface 3 43 ppm...
Страница 96: ......
Страница 108: ......
Страница 118: ......
Страница 122: ......
Страница 124: ......
Страница 129: ......