Operation
3.2
The AquaXact 1688 uses state-of-the-art microelectronics to provide a miniature full
functioning transmitter. The user should consider the following precautions when using any
sensitive electronic device:
Do not install the unit near heat sources such as radiators or air ducts.
Do not install the unit in places subject to extreme mechanical vibration
or shock.
Observe the appropriate electrical safety codes and regulations.
Electromagnetic compatibility considerations
3.3
The AquaXact 1688 has been designed and verified by testing to meet the requirements of
the EU EMC Directive 2014/30/EU, for Industrial, Scientific & Medical equipment. The
sensor and the 4-20 mA loop are electrically connected, however they are isolated from
the AquaXact 1688 housing and sample cell fitting threads. Please consider the following
electromagnetic interference issues during installation:
In order to provide an acceptable noise environment for the AquaXact 1688 or any
other digital equipment in the proximity of switching inductive loads, it is recommended
that there be varistors placed across the inductors to keep down the high voltage spikes
during transitions.
AC power wiring should be routed as far away from the AquaXact 1688 and its wiring
as practical.
Содержание AquaXact 1688
Страница 1: ...AquaXact 1688 Dew Point Transmitter Operator Manual...
Страница 15: ...AquaXact Issue 1 0 Servomex Group Limited 2017 15 Figure 4 4 Electrical installation connector and connections...
Страница 29: ...AquaXact Issue 1 0 Servomex Group Limited 2017 29 Appendix A Mechanical Drawings A 1 Dew point transmitter...
Страница 30: ...30 Servomex Group Limited 2017 AquaXact Issue 1 0 A 2 Sample block...
Страница 31: ...AquaXact Issue 1 0 Servomex Group Limited 2017 31 A 3 Sampling panel...
Страница 35: ...AquaXact Issue 1 0 Servomex Group Limited 2017 35 Appendix D Uncertainty in LBS ppmV calculations...