AquaXact Issue 1.0
© Servomex Group Limited 2017
19
7
Routine maintenance
The AquaXact 1688 Dew Point Transmitter does not contain any user
serviceable parts.
Do not attempt to maintain or service the AquaXact 1688 Dew Point
Transmitter unless you are trained and know what you are doing. The
transmitter must be maintained by a factory-trained field service
engineer.
Sample and calibration gases may be toxic or asphyxiant.
Never inspect the inlet filter(s), or service or repair the transmitter while
such gases are still connected to it.
If the transmitter is to be serviced or repaired it is important that all
pipework is flushed with an inert gas and the transmitter is allowed to
freely vent to local atmosphere.
Do not store replacement tips for more than 12 months before use to
assure the calibration is still valid.
Preventative maintenance
7.1
To minimize unscheduled instrument downtime, ensure the proper operation of the
AquaXact and to comply with the guidelines of applicable regulatory bodies, we
recommend that you utilize an annual preventative maintenance program for your
transmitter.
The preventative maintenance program consists of an annual inspection of the AquaXact,
and repair of any faults. To ensure that the AquaXact meets its original factory
specification we recommend that the sensor tip be replaced every year. This can be
accomplished by the user if equipped with a digital controller.
Contact Servomex or your local Servomex agent to arrange for a preventative
maintenance contract.
Содержание AquaXact 1688
Страница 1: ...AquaXact 1688 Dew Point Transmitter Operator Manual...
Страница 15: ...AquaXact Issue 1 0 Servomex Group Limited 2017 15 Figure 4 4 Electrical installation connector and connections...
Страница 29: ...AquaXact Issue 1 0 Servomex Group Limited 2017 29 Appendix A Mechanical Drawings A 1 Dew point transmitter...
Страница 30: ...30 Servomex Group Limited 2017 AquaXact Issue 1 0 A 2 Sample block...
Страница 31: ...AquaXact Issue 1 0 Servomex Group Limited 2017 31 A 3 Sampling panel...
Страница 35: ...AquaXact Issue 1 0 Servomex Group Limited 2017 35 Appendix D Uncertainty in LBS ppmV calculations...