generating a plot that will not be viewed. To resume plotting
data, simply uncheck the box.
Errors/Disconnection
Should there be a power interruption or momentary
disconnection, the LISST-ABS software will automatically
reconnect when the power or connection is reestablished. If the
disconnection occurs while logging data, the logging will pause
and resume automatically when the instrument is reconnected.
Should one of the two error codes be produced by the ABS, a
brief description of the problem will be shown. A value of -999
means low power supply voltage and the instrument cannot
operate. A value of -888 implies obstruction within sample
volume, or the second sample volume is at a boundary (see
Appendix C). In either case, this error is triggered when
backscatter from the second range cell is stronger than the first.