5.3 Calibration Wizard
For the user’s convenience a TTC/12TC-wizard is included in the software.
To start the wizard, please use the magic wand from the
Main Window
’s
toolbar or select
Extras
→
Three-Term Correction
from the main menu. A
selection dialog appears and lets the user choose between:
1.
Differential TTC (DTTC, available with differential measurement modes)
2.
Single-ended TTC (for channel 1 only, use DTTC for channel 2)
3.
Save current TTC/12TC for immediate reload
4.
Reload last known TTC/12TC
5.
Reset currently loaded TTC/12TC
6.
Full 2-port calibration (only if F2P module is available)
If you choose
, or
the TTC-wizard will guide you through the neces-
sary steps towards a reliable TTC, DTTC or 12TC; nevertheless subsequently
some helpful remarks are enlisted:
•
In the start window, an existing TTC/12TC can be reloaded. Please
ensure that the currently used measurement setup (in the first place
the attached cables and time base settings of the device) is identical
to that of the original TTC/12TC.
•
If a TTC/12TC is loaded from a previously
wizard-saved
file, the wiz-
ard jumps directly to the last window and you are done. This is only
necessary if different TTCs/12TCs (e.g. for different time bases) shall
be handled in parallel.
Please note:
This feature will become obsolete in future software releases. It
is recommended to save TTC/12TC calibrations as standard project mas-files
instead of wizard-saved sqc-files.