Power Management Feature
In the event of a Power Interruption:
Warning:
Because the duration of a power interruption is unknown, seek an alternate supply of oxygen
during the power interruption. Resume use of the Regalia when power is restored.
If a power outage occurs, the Regalia will activate both an audible and a visual red light alarm to inform you that the
Regalia is not operating. The Regalia should be located where the alarms can be seen and heard.
If the duration of the interruption is shorter than twenty seconds and the unit has not been turned off, the Regalia will
resume operation twenty seconds after the initial interruption.
Important:
During the power interruption:
The user can press the ON/OFF button to silence the alarms and turn the unit off.
If the unit
was not
turned off the Regalia will automatically resume operation once power is restored.
If the unit
was
turned off during the power interruption, the user must press the ON/OFF button again, after
the power is restored, to turn the unit on.
Operating Conditions
1.
Control Panel Display –
If the
concentrator is an OM Model (equipped with an oxygen monitor), the
oxygen concentration is being continuously monitored while the Regalia is running and oxygen is being
produced.
If the concentration level falls below preset levels, an indicator light will activate. Operating
conditions are described in the table on the following page.
Green Light
On/Off Button
Red Light
Yellow Light
Increase Flow Button
Decrease Flow Button
LCD Flow Indicator
(LPM)