Maintenance 

 

Routine Maintenance: 

Routine maintenance consists of cleaning the air inlet filter at regular intervals to keep 

the unit functioning for years.  The user can easily achieve this maintenance.  
 

Warning: 

Unplug the oxygen concentrator from the power outlet before cleaning or routine 

maintenance. Use a damp cloth for cleaning. 

Do Not

 

directly apply

 

liquid spray cleaners or 

aerosol cleaners. 
 
 

 

Cabinet and Power Cord 

 

The cabinet and power cord can be cleaned with a mild detergent solution applied with a damp cloth and 
towel dried. 

 
 

 

Air Inlet Filter 

 

Room air is drawn into the oxygen concentrator through the air inlet filter located on the top backside of the 
cabinet.  Cleaning the air inlet filter is the most important activity you will perform to keep your oxygen 
concentrator performing properly.  Check this air filter daily and clean it at least once per week. 

 

1.

 

Remove the filter from the cabinet. 

 

2.

 

Wash the filter in warm water using a mild detergent solution. 

 

3.

 

Rinse the filter thoroughly and squeeze out the excess water.

 

 

4.

 

Allow the filter to air dry.  

 

 

Caution:

 The filter should be free of liquid water and dry to the touch before reinstallation, as 

excess moisture may impair the proper operation of the oxygen concentrator. 

 

5.

 

Reinsert the filter in the cabinet. 

 

Caution: 

The air inlet filter should be replaced as needed.  If the oxygen concentrator is used in 

a dusty environment, the filter may need to be replaced more often. 

Do Not

 

operate the oxygen 

concentrator for more than 30 minutes without a filter installed. 

 
 

 

Oxygen Delivery Accessories - 

Follow the instructions supplied by the manufacturer.  

 

 

 

Содержание Regalia

Страница 1: ...Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P N 5167 Rev A March 2006 ...

Страница 2: ...nment Guidelines 6 The Proper Location 6 Before Operating Your Oxygen Concentrator 6 Operating Your Oxygen Concentrator 7 Installation Instructions 8 Select a Location 8 Inspect Inlet Air Filter 8 On Off Button 8 Warm up Period 8 Flow Rate Adjustment 8 Power Management Feature 9 Operating Conditions 9 Control Panel Display 9 Display of Operating Conditions 10 Non Usage 10 Maintenance 11 Trouble Sh...

Страница 3: ...void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen concentrator tubing and connections and any other oxygen source Do Not use any petroleum based or other lubricants unless recommended by SeQual Technologies Inc Keep oxygen concentrato...

Страница 4: ...ng information specific instructions commentary sidelights or interesting points Symbols Used in Instruction Manual and on Oxygen Concentrator Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided ...

Страница 5: ...of protection against electrical shock than provided by type B applied part Class II Equipment Equipment in which protection against electrical shock does not rely on basic insulation only but in which additional safety precautions are provided Amperes Alternating Current Oxygen Concentrator Information SeQual s Regalia Oxygen Concentrators are quiet reliable and low maintenance sources of oxygen ...

Страница 6: ...ake filter on the top of the cabinet without being restricted Make sure the oxygen concentrator is at least three to six 3 6 inches 7 6 15 2 cm away from walls furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen concentrator Do Not cover unit with a blanket towel quilt etc Do Not place objects under unit Maintain at least inch clearance under unit...

Страница 7: ...ference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer or authorized service representative General Precautions Retain the product s safety and operating instructions for future reference Follow all operating and usage instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily...

Страница 8: ... will illuminate to indicate the unit is running 4 Warm up Period After initially turning on the unit allow three 3 to ten 10 minutes for the unit to reach its specified performance If the unit is equipped with an oxygen concentration monitor OM models only the red yellow and green light will intermittently illuminate while the unit is warming up For details on warm up refer to the Display of Oper...

Страница 9: ...tion Important During the power interruption The user can press the ON OFF button to silence the alarms and turn the unit off If the unit was not turned off the Regalia will automatically resume operation once power is restored If the unit was turned off during the power interruption the user must press the ON OFF button again after the power is restored to turn the unit on Operating Conditions 1 ...

Страница 10: ...n Off On Quickly Yes Off Flashing Off Off Running Oxygen Concentration Below Normal Level Yes Off Off Flashing On Running Oxygen Concentration Abnormal Level 1 Yes Off On Off Off Running Button Stuck Yes On On Off Off Running 9V Battery Low Voltage if enabled No Off Off Flashing On Not Running Unit is overheating Compressor Malfunction Power Interruption Internal Fuse Blown No On or Off On or Off ...

Страница 11: ...e air inlet filter is the most important activity you will perform to keep your oxygen concentrator performing properly Check this air filter daily and clean it at least once per week 1 Remove the filter from the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry Caution The filter sho...

Страница 12: ...ack Light On Unit overheating Remove any obstruction s from air openings and refer to page 6 The Proper Location Compressor malfunction Service required Table 2 Concentrator Running Symptom Probable Cause Remedy No oxygen flow see Alarm Indications in Table on Page 11 Tubing blocked or kinked Clean or replace tubing Remove tube kink Air Inlet Filter blocked Clean air inlet filter Internal componen...

Страница 13: ...ion Excessive Internal Temperature Outlet Pressure 35 36 psig 241 2 248 1 kPa relief valve Low Oxygen Output 70 oxygen Back up Alarm Power 9V Battery Filters Cabinet Air Inlet Compressor Inlet and HEPA Device Classification IEC Class II Type BF Applied Part Continuous Operation Transport Storage Requirements Temperature 4 F 20 C to 140 F 60 C Humidity Up to 95 Non Condensing Back pressure Effect o...

Страница 14: ...ul interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation distance between the equipment Connect the equipment into an outlet on a circuit different from that which the other device s are connected Co...

Отзывы: