Recommended Operating Environment Guidelines  

 

  Temperature 

  50

º

 F to 104

º

 F (10

º

 C to 40

º

 C) 

  Electrical 

  Use no extension cords or electrical outlets controlled by a switch. 

  Placement 

  No closer than 3 inches (7.6 cm) from the wall, draperies and furniture. 

  Environment 

  Smoke, pollutant, and fume free.  Relative Humidity: 10% - 95% 

  Operating Time 

  Up to 24 hours per day. 

  Flow Rate 

  From 0.5 LPM to 10.0 LPM (for the 10 LPM version) 

  From 0.5 LPM to 7.0 LPM (for the 7 LPM version) 

                                          

 

 0

 

  0

 

30

 

35

 

40

 

55

 

60

 

65

 

70

 

75

 

80

 

85

 

90

 

95

 

Relative Humidity, %

 

Te

m

p

e

ra

tu

re

°C

 

 

 

Important: 

For rated performance at 10 LPM (or 7 LPM for the 7 LPM version) do not operate the Oxygen 

concentrator above the indicated boundaries.  Operating at flow rates less than 10 (or 7 LPM for the 7 LPM 
version) LPM will tolerate higher temperature/humidity conditions. 

 
 
 

 

The Proper Location 

 

Usage in environments other than those described on the previous page may result in the need for increased 
oxygen concentrator maintenance.  Select a location in a well ventilated space that avoids fumes and pollutants 
and allows the oxygen concentrator to draw in room air through the intake filter on the top of the cabinet without 
being restricted. Make sure the oxygen concentrator is at least three to six (3 - 6) inches (7.6 – 15.2 cm) away 
from walls, furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen 
concentrator.  

Do Not

 cover unit with a blanket, towel, quilt, etc. 

Do Not

 place objects under unit. Maintain at 

least ½ inch clearance under unit.  Keep the oxygen concentrator at least five (5) feet (1.5 m) away from hot, 
sparking objects or open sources of flame.  

Do Not

 

locate the unit near any heat source or heat register. 

 
Before Operating Your Oxygen Concentrator 

 

There are many different types of oxygen tubing that can be used with this oxygen concentrator.   

 

 

Oxygen Tubing - C

onnect the oxygen tubing directly to the oxygen outlet port. Connect the other end of the 

tube to the application, if not already attached.  

Important:

 

The tube fitting may be tight

 - 

Do Not

 

use oil or 

grease to lubricate

 To ensure there is enough pressure to deliver oxygen, the length of the tubing should not 

exceed 50 feet (15.2m). Maximum extension tubing length is dependent upon environmental 
conditions, tubing type, liter flow and any accessories used between DISS fitting and 
application. Shorter tubing lengths will be required in some cases.  The internal diameter of 
the tubing should be no less than 3/16-inch (0.48 cm).   

 

Содержание Regalia

Страница 1: ...Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P N 5167 Rev A March 2006 ...

Страница 2: ...nment Guidelines 6 The Proper Location 6 Before Operating Your Oxygen Concentrator 6 Operating Your Oxygen Concentrator 7 Installation Instructions 8 Select a Location 8 Inspect Inlet Air Filter 8 On Off Button 8 Warm up Period 8 Flow Rate Adjustment 8 Power Management Feature 9 Operating Conditions 9 Control Panel Display 9 Display of Operating Conditions 10 Non Usage 10 Maintenance 11 Trouble Sh...

Страница 3: ...void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen concentrator tubing and connections and any other oxygen source Do Not use any petroleum based or other lubricants unless recommended by SeQual Technologies Inc Keep oxygen concentrato...

Страница 4: ...ng information specific instructions commentary sidelights or interesting points Symbols Used in Instruction Manual and on Oxygen Concentrator Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided ...

Страница 5: ...of protection against electrical shock than provided by type B applied part Class II Equipment Equipment in which protection against electrical shock does not rely on basic insulation only but in which additional safety precautions are provided Amperes Alternating Current Oxygen Concentrator Information SeQual s Regalia Oxygen Concentrators are quiet reliable and low maintenance sources of oxygen ...

Страница 6: ...ake filter on the top of the cabinet without being restricted Make sure the oxygen concentrator is at least three to six 3 6 inches 7 6 15 2 cm away from walls furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen concentrator Do Not cover unit with a blanket towel quilt etc Do Not place objects under unit Maintain at least inch clearance under unit...

Страница 7: ...ference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer or authorized service representative General Precautions Retain the product s safety and operating instructions for future reference Follow all operating and usage instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily...

Страница 8: ... will illuminate to indicate the unit is running 4 Warm up Period After initially turning on the unit allow three 3 to ten 10 minutes for the unit to reach its specified performance If the unit is equipped with an oxygen concentration monitor OM models only the red yellow and green light will intermittently illuminate while the unit is warming up For details on warm up refer to the Display of Oper...

Страница 9: ...tion Important During the power interruption The user can press the ON OFF button to silence the alarms and turn the unit off If the unit was not turned off the Regalia will automatically resume operation once power is restored If the unit was turned off during the power interruption the user must press the ON OFF button again after the power is restored to turn the unit on Operating Conditions 1 ...

Страница 10: ...n Off On Quickly Yes Off Flashing Off Off Running Oxygen Concentration Below Normal Level Yes Off Off Flashing On Running Oxygen Concentration Abnormal Level 1 Yes Off On Off Off Running Button Stuck Yes On On Off Off Running 9V Battery Low Voltage if enabled No Off Off Flashing On Not Running Unit is overheating Compressor Malfunction Power Interruption Internal Fuse Blown No On or Off On or Off ...

Страница 11: ...e air inlet filter is the most important activity you will perform to keep your oxygen concentrator performing properly Check this air filter daily and clean it at least once per week 1 Remove the filter from the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry Caution The filter sho...

Страница 12: ...ack Light On Unit overheating Remove any obstruction s from air openings and refer to page 6 The Proper Location Compressor malfunction Service required Table 2 Concentrator Running Symptom Probable Cause Remedy No oxygen flow see Alarm Indications in Table on Page 11 Tubing blocked or kinked Clean or replace tubing Remove tube kink Air Inlet Filter blocked Clean air inlet filter Internal componen...

Страница 13: ...ion Excessive Internal Temperature Outlet Pressure 35 36 psig 241 2 248 1 kPa relief valve Low Oxygen Output 70 oxygen Back up Alarm Power 9V Battery Filters Cabinet Air Inlet Compressor Inlet and HEPA Device Classification IEC Class II Type BF Applied Part Continuous Operation Transport Storage Requirements Temperature 4 F 20 C to 140 F 60 C Humidity Up to 95 Non Condensing Back pressure Effect o...

Страница 14: ...ul interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation distance between the equipment Connect the equipment into an outlet on a circuit different from that which the other device s are connected Co...

Отзывы: