SeQual Regalia Скачать руководство пользователя страница 4

 

 

SeQual Technologies Inc.    

  

Part No: 5167 

Indications for Use 
 

The SeQual Regalia Oxygen Concentrator is intended to provide only non-medical oxygen. It is not for medical 
purposes.  

 

This oxygen concentrator is not intended for life supporting or life sustaining applications. 

 
 

Introduction 

 

SeQual's Regalia Oxygen Concentrator is an electronically operated oxygen concentrator that separates oxygen 
from room air.  It does not remove oxygen from a room any more than by simply breathing the same room air. 
You utilize the oxygen by attaching tubing to the oxygen outlet fitting. 

 

 
As  with  any  electrically  powered  device,  the  user  may  experience  periods  of  non-operation  as  the  result  of 
electrical  power  interruption,  or  the  need  to  have  the  oxygen  concentrator  serviced  by  a  technician.    This 
instruction manual will inform you about the use and care of the Regalia Oxygen Concentrator and will serve as a 
reference as you use your oxygen concentrator. 

 

 

Important Safety Instructions are shown under the following headings: 

 

Warning:

  

Important safety information for hazards that might cause serious injury. 

Caution:

  

Information for preventing damage to the oxygen concentrator.

 

Do Not:  

Information to which you should pay special attention. 

Important:

 

 Clarifying information, specific instructions, commentary, sidelights, or interesting points.  

 

Symbols Used in Instruction Manual and on Oxygen Concentrator 

 

Refer to this section for an explanation of the symbols and the warnings that accompany them. 
 

"Read Instructions"

  This symbol, when used alone or in conjunction with any of the following 

symbols, indicates the need to consult the operating instructions provided with the product. A 
potential risk exists if the operating instructions are not followed. 

"Attention or Caution"

  Consult accompanying documents.  This symbol indicates important 

information that requires the user to take special precautions when certain conditions are present.   

"No Smoking or Naked Flames" 

The use of oxygen requires that special care be taken to reduce 

the risk of fire.  Materials that burn in air and even some material that will not burn in air are 
easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, all possible 
sources of ignition must be kept away from the oxygen concentrator and preferably out of the 
room in which it is being used.

 

Smoking in the proximity of an operating oxygen concentrator is 

dangerous, and can permanently damage the device and void the warranty. 

"Use No Oil or Grease"

 Indicates important information about the possibility of spontaneous 

and violent ignition that may occur if oil, grease or other petroleum based substances come into 
contact with oxygen under pressure. 

Содержание Regalia

Страница 1: ...Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P N 5167 Rev A March 2006 ...

Страница 2: ...nment Guidelines 6 The Proper Location 6 Before Operating Your Oxygen Concentrator 6 Operating Your Oxygen Concentrator 7 Installation Instructions 8 Select a Location 8 Inspect Inlet Air Filter 8 On Off Button 8 Warm up Period 8 Flow Rate Adjustment 8 Power Management Feature 9 Operating Conditions 9 Control Panel Display 9 Display of Operating Conditions 10 Non Usage 10 Maintenance 11 Trouble Sh...

Страница 3: ...void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen concentrator tubing and connections and any other oxygen source Do Not use any petroleum based or other lubricants unless recommended by SeQual Technologies Inc Keep oxygen concentrato...

Страница 4: ...ng information specific instructions commentary sidelights or interesting points Symbols Used in Instruction Manual and on Oxygen Concentrator Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided ...

Страница 5: ...of protection against electrical shock than provided by type B applied part Class II Equipment Equipment in which protection against electrical shock does not rely on basic insulation only but in which additional safety precautions are provided Amperes Alternating Current Oxygen Concentrator Information SeQual s Regalia Oxygen Concentrators are quiet reliable and low maintenance sources of oxygen ...

Страница 6: ...ake filter on the top of the cabinet without being restricted Make sure the oxygen concentrator is at least three to six 3 6 inches 7 6 15 2 cm away from walls furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen concentrator Do Not cover unit with a blanket towel quilt etc Do Not place objects under unit Maintain at least inch clearance under unit...

Страница 7: ...ference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer or authorized service representative General Precautions Retain the product s safety and operating instructions for future reference Follow all operating and usage instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily...

Страница 8: ... will illuminate to indicate the unit is running 4 Warm up Period After initially turning on the unit allow three 3 to ten 10 minutes for the unit to reach its specified performance If the unit is equipped with an oxygen concentration monitor OM models only the red yellow and green light will intermittently illuminate while the unit is warming up For details on warm up refer to the Display of Oper...

Страница 9: ...tion Important During the power interruption The user can press the ON OFF button to silence the alarms and turn the unit off If the unit was not turned off the Regalia will automatically resume operation once power is restored If the unit was turned off during the power interruption the user must press the ON OFF button again after the power is restored to turn the unit on Operating Conditions 1 ...

Страница 10: ...n Off On Quickly Yes Off Flashing Off Off Running Oxygen Concentration Below Normal Level Yes Off Off Flashing On Running Oxygen Concentration Abnormal Level 1 Yes Off On Off Off Running Button Stuck Yes On On Off Off Running 9V Battery Low Voltage if enabled No Off Off Flashing On Not Running Unit is overheating Compressor Malfunction Power Interruption Internal Fuse Blown No On or Off On or Off ...

Страница 11: ...e air inlet filter is the most important activity you will perform to keep your oxygen concentrator performing properly Check this air filter daily and clean it at least once per week 1 Remove the filter from the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry Caution The filter sho...

Страница 12: ...ack Light On Unit overheating Remove any obstruction s from air openings and refer to page 6 The Proper Location Compressor malfunction Service required Table 2 Concentrator Running Symptom Probable Cause Remedy No oxygen flow see Alarm Indications in Table on Page 11 Tubing blocked or kinked Clean or replace tubing Remove tube kink Air Inlet Filter blocked Clean air inlet filter Internal componen...

Страница 13: ...ion Excessive Internal Temperature Outlet Pressure 35 36 psig 241 2 248 1 kPa relief valve Low Oxygen Output 70 oxygen Back up Alarm Power 9V Battery Filters Cabinet Air Inlet Compressor Inlet and HEPA Device Classification IEC Class II Type BF Applied Part Continuous Operation Transport Storage Requirements Temperature 4 F 20 C to 140 F 60 C Humidity Up to 95 Non Condensing Back pressure Effect o...

Страница 14: ...ul interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation distance between the equipment Connect the equipment into an outlet on a circuit different from that which the other device s are connected Co...