activated only every certain number or count of shots in the flash lamp. Then, the control of the
flash lamp would be taken directly from "Out A" while the active Q-switch would be controlled
from "Out B" and a third external component, for example an optical spectrometer, would be
controlled from the output "Out C " For the input configuration menu and to activate the counter,
you can consult page 12 of this manual.
The following figure shows the behavior of the outputs when the retarder is configured with
delays "dA", "dB", "dC" and pulse widths "Wa", "Wb" and "Wc" respectively on the output
channels "Out A ”,“ Out B ”and“ Out C ”, while the“ IN ”input is configured to use the rising edge of
an input pulse as“ zero ”time or reference to start the delays of each output. During the emission
of a pulse in some output channel, said channel will not attend to another event in the input
channel, and this does not affect neighboring channels.
Fig. 4. Behavior of the outputs with respect to an input pulse, pulse counter deactivated