20
Sentry Equipment Corp
Clean
Figure 18. VREL Cleaning Frequency
A u t o
C l e a n
F r e q u e n c y
d d : h h : m m
0 : 0 0 : 0 0
Figure 18 shows the Auto Cleaning Frequency screen. This screen sets the frequency
between auto cleaning cycles for the VREL valve. When auto clean is enabled the
controller periodically actuates the VREL valve to prevent plugging. To change the minutes
field, use the
▲
and
▼
keys. To change the hours field, hold the ENTER key while using
the
▲
and
▼
keys. To change the days field, hold the MENU key while using the
▲
and
▼
keys. To return to the continued setup screen, use the ESC key.
Â
NOTE
To disable auto cleaning, set the frequency to 0:00:00 .
Continued Setup (Screen 2)
Figure 19. Continued Setup (Screen 2)
S e t u p
C o n t i n u e d
►
A l a r m s
M O D B U S
U n i t s
P I D
O p t i o n s
Figure 19 shows the second continued setup screen. The cursor indicates the option you
wish to choose. Move the cursor using the
▲
and
▼
keys. Select the option using the
ENTER key. To exit this screen and return to the previous screen, press the ESC key.
Alarm Setup
Figure 20. Alarm Setup
A l a r m
S e t u p
O u t l e t
T e m p e r a t u r e
1 2 0
F
Figure 20 shows the alarm setup screen. Use the
▲
and
▼
keys to change the alarm
threshold. Use the ENTER key to advance to the next alarm point. After all alarm levels
are set, the screen automatically returns to the continued setup screen (screen 2). The
following alarm conditions can be set:
Outlet Temperature:
High temperature alarm for the sample temperature to analyzers.
A high outlet temperature triggers a shutdown and sets the common alarm relay. On
later model units (11/16/11 or later) the high temperature setpoint can be changed when
in a high temperature alarm by holding the ESC button down and pressing the Enter
button. This sets the high temperature setpoint to 5 degrees above the current outlet
Содержание Sample Sentry II High Pressure
Страница 7: ...Sentry Sample Sentry II 7 Figure 1 Typical P ID of a Sample Sentry II System...
Страница 8: ...8 Sentry Equipment Corp Figure 2 P ID of a Sample Sentry II system with Secondary Cooling...
Страница 11: ...Sentry Sample Sentry II 11 Figure 3 Controller Front Panel...
Страница 24: ...24 Sentry Equipment Corp Figure 25 General Arrangement High Pressure Model...
Страница 26: ...26 Sentry Equipment Corp Figure 26 Electrical Wiring Diagram...
Страница 43: ...Sentry Sample Sentry II 43 This page is intentionally left blank...
Страница 44: ...44 Sentry Equipment Corp This page is intentionally left blank...