Sentry Sample Sentry II 13

Operation Screen

The operation screen is the main screen on the Sample Sentry II. This screen contains 

information indicating the status of the system. While in Manual mode, the system is 

controlled from the operation screen.

Figure 5. Operation Screen

I n l e t

1 5 0 0

P S I

B l o w d o w n

1

4

0

7

c c

0 : 2 7 : 3 7

m i n

S V 1

S V 2

P R V

Figure 5 shows a typical operation screen for the Sample Sentry II. The top line displays 

temperature and pressure data. The screen periodically refreshes to show a different 

measurement. Only measurements that are enabled are displayed.

Line 4 displays the state of the three valves. Figure 5 currently shows valve SV1 open, SV2 

closed, and the PRV partially open. If the PRV display was completely filled in, it would 

indicate that the PRV was fully closed.

Line 2 displays the state of the system, including the total flow rate. The 7 possible states 

are 

Manual

Start

Blowdown

Sample

Stop

Remote

, and 

Clean

Manual:

 While in manual mode, the system can be controlled with the keypad on the 

front of the controller. Pressing the MENU key displays the Main Menu password screen. 

To place the system in Automatic mode, press the ENTER key. Pressing the ESC key toggles 

valves V1 and SV2 between open and closed. Holding the 

 key opens the PRV, while 

holding the 

 key closes the PRV.

Start:

 The controller waits in the Start state until the inlet pressure is greater than the 

programmed startup pressure. The only key that is functional while in the Start state is the 

ENTER key. Pressing the ENTER key returns the system to 

Manual

 mode.

Blowdown:

 Flow to analyzers is stopped by closing SV2, forcing the entire sample to flow 

to the grab sample. When the system is in 

Blowdown

, the time remaining in Blowdown is 

displayed on line 3. Pressing the ENTER key returns the system to 

Manual

 mode. Pressing 

the ESC key advances the system to 

Sample

.

Sample:

 The controller opens SV2, allowing flow to analyzers. Pressing the ENTER key 

returns the system to 

Manual

 mode. Pressing the ESC key returns the system to 

Blowdown

.

Stop:

 The Stop state occurs when the total flow falls below a shutdown set point and the 

PRV is full open or if the pressure is lower than the shutdown pressure set point. During 

the Stop state V1 is closed, the PRV is adjusted to its set point, and SV2 is closed. After 

these valves are adjusted, the controller advances to 

Start

 if automatic startup is enabled 

or 

Manual

 if disabled.

Remote:

 Remote indicates the system is being controlled remotely with the MODBUS 

communications port. While in the Remote mode, the controller enters the Blowdown and 

Sample states as determined from remote commands. Pressing the ENTER key returns the 

system to Manual mode.

Содержание Sample Sentry II High Pressure

Страница 1: ...Original Instructions Installation Operation Maintenance Manual Sentry Sample Sentry II Sample Conditioning Automation S SW IOM 00269 20 05 21...

Страница 2: ...abilities and warranties express or implied including warranties of merchantability and fitness for a particular purpose for the accuracy currency completeness and or reliability of the information co...

Страница 3: ...ions DANGER indicates a hazardous situation which if not avoided will result in death or serious injury WARNING indicates a hazardous situation which if not avoided could result in death or serious in...

Страница 4: ...g with any chemical Read the product label and Safety Data Sheets SDS carefully and follow the instructions exactly Potential Equipment Hazards Hot surfaces This equipment may have very hot surfaces I...

Страница 5: ...Flow Rate Pressure Measurements Range 0 3000 psi 0 207 bar Accuracy 0 5 Temperature Measurements Range 32 250 F 0 121 C Accuracy 1 F Alarm Relay Contact Rating Max operating voltage 250 VAC 220 VDC M...

Страница 6: ...m can require constant adjustment during startup or as plant conditions change Sample Sentry II is designed to meet requirements for representative sampling by maintaining a constant flow rate with va...

Страница 7: ...Sentry Sample Sentry II 7 Figure 1 Typical P ID of a Sample Sentry II System...

Страница 8: ...8 Sentry Equipment Corp Figure 2 P ID of a Sample Sentry II system with Secondary Cooling...

Страница 9: ...nt the system advances to the Blowdown state If Blowdown is not enabled the system goes to the Sample state When enabled the Blowdown state initiates a controlled blowdown sequence by opening V 1 and...

Страница 10: ...n for the main menu In order to enter the menu screen you must enter the password 101 followed by the ENTER key If a correct password is entered the main menu screen appears From an alarm screen press...

Страница 11: ...Sentry Sample Sentry II 11 Figure 3 Controller Front Panel...

Страница 12: ...Scale Sample Auto Calibra on Shutdown Manual Cal Blowdown Time Comp Error Offset Pressures Startup Pressure Flow Meter Cal Shutdown Pressure PRV Se ngs PRV Type Clean PRV Start Posi on PRV Recalibrate...

Страница 13: ...nd closed Holding the key opens the PRV while holding the key closes the PRV Start The controller waits in the Start state until the inlet pressure is greater than the programmed startup pressure The...

Страница 14: ...rm occurs valves SV1 SV2 and the VREL valve close The alarm condition cannot be reset until the outlet temperature falls below the high temperature alarm setpoint Pressing the Menu Reset button resets...

Страница 15: ...was in before Motor Diagnostics routine 3 If 15 seconds have elapsed and no motor movement has been detected the Sample Sentry II pulses the motor in the opposite direction for 15 seconds If motor mov...

Страница 16: ...he option by using the ENTER key To exit this screen and return to the operation screen press the ESC key The Main Menu password is 101 Display Intensity Figure 8 Display Intensity D i s p l a y I n t...

Страница 17: ...I recommends a blowdown velocity no greater than 125 the normal sampling velocity Sample This flow rate is used during the Sample state During this time sample is allowed to flow to analyzers Shutdown...

Страница 18: ...s the ENTER key to display the next set point After setting the last pressure pressing ENTER returns the unit to the setup screen The following pressures can be programmed Startup The first set point...

Страница 19: ...position fully closed This can be set to higher if the sample line tends to have initial debris in the line that can clog the VREL valve or to reduce the time it takes for the sample to reach its set...

Страница 20: ...d continued setup screen The cursor indicates the option you wish to choose Move the cursor using the and keys Select the option using the ENTER key To exit this screen and return to the previous scre...

Страница 21: ...ystem remains in low pressure alarm until the pressure rises above the low alarm level plus the alarm hysteresis Units Figure 21 Units U n i t s S e t u p E n g l i s h Figure 21 shows the unit select...

Страница 22: ...e 24 shows the analog input menu screen This screen can be accessed by entering 202 for the Analog Input Setup Password The cursor indicates the option you wish to choose Move the cursor using the and...

Страница 23: ...n This instrument should be installed by qualified technical personnel to ensure adherence to all applicable electrical codes To ensure the protection provided by this equipment is not impaired this e...

Страница 24: ...24 Sentry Equipment Corp Figure 25 General Arrangement High Pressure Model...

Страница 25: ...A suitable external over current protection device such as a fuse or circuit breaker 15 A is recommended for both the hot L and neutral N leads Using this method the plug serves as the disconnect devi...

Страница 26: ...26 Sentry Equipment Corp Figure 26 Electrical Wiring Diagram...

Страница 27: ...tion valve to introduce sample The controller begins to display the inlet pressure 6 Place the controller in Auto mode for automatic operation or control the system manually Maintenance Sample Cooler...

Страница 28: ...ces startup time a When the coupling is installed it is necessary to ensure proper spacing See procedure in maintenance section 6 Apply power to the Sample Sentry II 7 Wait for Sample Sentry II unit t...

Страница 29: ...ator assembly by removing the two tubing connectors and two Allen head retaining screws that hold the valve to the mounting bracket Flow Meter FE FT 1 If removal of the flow transmitter is needed spec...

Страница 30: ...ween the top and bottom of the coupling when pulling the VREL valve shaft up e Tighten the setscrews and install tighten the jam nuts Low Pressure Needle Valve coupling 1 Install the bronze half of th...

Страница 31: ...cover of the enclosure by removing the four corner screws Disconnect the keypad extension cable from the CPU board 2 Disconnect the interface cable and power supply cable from the CPU board Remove the...

Страница 32: ...ed personnel Refer to the following procedures to calibrate the analog circuits of the Sample Sentry II Tools Required Calibrated milliamp supply Calibrated voltage source Calibration 1 Using the keyp...

Страница 33: ...of channels 1 3 connect a calibrated milliamp supply to TB1 terminals of the CPU board marked and for the appropriate channel Input 4 mA Press the Enter key 4 The controller is done calibrating the 4...

Страница 34: ...ges from 0 V to 2 5 V Input 2 5 V for selected channel Press the Enter key 10 The controller is done calibrating the 2 5 V when the Gain value stops adjusting Remove voltage from channel Press the Ent...

Страница 35: ...n 0 cc min 1250 cc min 1000 cc min 0 cc min Blowdown Time 1 minute 1 minute Pressures Startup Shutdown 1000 PSI 500 PSI 100 25 Auto clean 8 hours 8 hours Alarms Outlet Temp Low Flow Low Flow Hysteresi...

Страница 36: ...t or replace packing Leaking Cooler Loose housing from thermal shock or worn gasket Tighten bolts on head assembly or replace gasket if cut or worn Leaking BPR RV Diaphragm failure Replace diaphragm M...

Страница 37: ...4281E V 2 Sample Inlet Isolation Blowdown Valves 1 4 Compression Low Temperature Low Pressure Packing for above 1 4 Compression High Temperature High Pressure Packing for above 3 8 Compression Low Tem...

Страница 38: ...H F1 CPU Board Fuse 1 0 A 250 V Fast Acting 4 05630B F1 Output Board Fuse 4 0 A 250 V Fast Acting 4 05630C F2 Output Board Fuse 0 63 A 250 V Fast Acting 4 05630D F3 Output Board Fuse 0 25 A 250 V Fast...

Страница 39: ...in the table below Product Line Product Category Warranty Period Sentry 1 Automatic Sampling 2 Corrosion Monitoring 3 Manual Sampling 4 Sample Conditioning 5 Sampling Analysis Systems 6 Replacement Pa...

Страница 40: ...nditioning sampling and measuring gas liquid slurry powder solids steam or water within their production environments our customers obtain the critical insights they need to control and optimize their...

Страница 41: ...mmunications port uses 8 data bits 1 stop bit and no parity The MODBUS mode is RTU These parameters cannot be changed Table 6 MODBUS Registers Register Sample Sentry II Register R W Format 30001 State...

Страница 42: ...r back to zero Remote Stop Permissive Run 40017 Momentarily place a 1 in this register to remotely place the system in Manual Maintaining a 1 in this register functions as a remote lockout To place a...

Страница 43: ...Sentry Sample Sentry II 43 This page is intentionally left blank...

Страница 44: ...44 Sentry Equipment Corp This page is intentionally left blank...

Страница 45: ...sentry equip com 966 Blue Ribbon Circle North Oconomowoc WI 53066 U S A 1 262 567 7256 support sentry equip com Serving customers in more than 50 countries across six continents worldwide...

Отзывы:

Похожие инструкции для Sample Sentry II High Pressure