ST801B Optical Power Multi-Meter 

 

damage eyes) 

OPM Connector: 

optical power test interface 

 

USB  Port:

  the  instrument  communicates  with  the 

management software by USB port 

Charge  hole:

  connect  the  instrument  with  the  charger  for 

power charging. 

5. Operations 

On/Off key:

   

Pressing 

“ ”

 for over 3 seconds, the tester can open. 

When the instrument is power-on, it defaults  optical power 

meter  module  and  1310nm  wavelength.  LCD  shows  the 

working  wavelength,  optical  power,  battery  power, 

modulation frequency identification and so on. Pressing 

“ ”

 

for over 3 seconds when tester is open, the tester will shut 

down.  System  defaults  automatic  power  off  in  10  minutes, 

the LCD will show

 

Auto-OFF

 

Open and close auto-off function:

   

In  optical  power  meter  module,  shortly  pressing 

“ ”

auto-off  function  logo  will  disappear  in  LCD.  It  means 

Содержание ST801B

Страница 1: ...ST801B Optical Power Multi Meter User Manual V1 0 Shandong Senter Electronic Co Ltd ...

Страница 2: ...ST801B Optical Power Multi Meter 2 Contents 1 Summary 3 2 Functions 4 3 Technical Specifications 5 4 Layout 7 5 Operations 9 6 Maintenance 12 7 Faults Solutions 13 ...

Страница 3: ...y advances the test efficiency The inner microprocessor and linearity magnification technology ensure the long time calibration and test accuracy Optical Power Multi meter can carry out not only the absolute and relative power test but also the optical loss test for fiber optical cable and optical passive devices automatically and it can save test results It can be widely used in LAN WLAN MAN CATV...

Страница 4: ...Hz and auto wavelength identification mode 2 8 Wavelength identification with the help of light modulation and identification mode 2 9 USB communication function can upload the test result to management software by USB port 2 10 Handheld large LCD display easy to operate 2 11 Real time clock function can real time display the current time provide recording time for data storage 2 12 Auto power off...

Страница 5: ...1dBm 6 Automatic frequency identification 330Hz 270Hz 1 KHz 2 KHz 7 Automatic wavelength identification with the help of corresponding modulation mode Optical source module 1 Emitter type FP LD DFB LD 2 Dual wavelength switch function 1310 20nm 1550 20nm 3 Spectral line width 5nm 4 Output optical power 7 dBm 0 dBm customized 5 Short time stability of output optical power 0 05dB 15min 6 Long time s...

Страница 6: ...itor inner battery voltage LCD display 3 Auto power off 10 minutes 4 1000 groups of test data storage and management 5 Battery 7 4V 1000mAh high performance rechargeable lithium battery or 4pcs AA dry battery 6 Working duration for battery 90h Li battery only optical power meter work 24h Li battery the optical source and optical power meter work synchronously 7 Operating temperature 10 40 8 Storag...

Страница 7: ...ST801B Optical Power Multi Meter 7 4 Layout ...

Страница 8: ...T801B Optical Power Multi Meter 8 LD Connector Light Source output interface When the light is turned on please don t look directly into optical output or connected fiber optic connectors in order not to ...

Страница 9: ...ster can open When the instrument is power on it defaults optical power meter module and 1310nm wavelength LCD shows the working wavelength optical power battery power modulation frequency identification and so on Pressing for over 3 seconds when tester is open the tester will shut down System defaults automatic power off in 10 minutes the LCD will show Auto OFF Open and close auto off function In...

Страница 10: ... 1550nm and 1625nm through this key The chosen wavelength will be displayed in LCD screen When the instrument is power on it defaults the wavelength of 1310nm and circulates in order of 1310nm 1490nm 1550nm 1625nm 850nm 1300nm 2 Show absolute power and relative power value The instrument defaults the absolute measurement of optical power The value displayed by main display 2 means logarithmic powe...

Страница 11: ...e time when there is some working light source And it is forbidden to look at the output end of laser source or the connector of the optical fiber when the light source is under the power on status The LCD displays the corresponding wavelength and modulation status when the tester is under the module of stable light source When it is back to the module of optical power the LCD displays the informa...

Страница 12: ...look directly at the optical output or the connected fiber optic connectors when the light turned on so as not to damage the eyes NOTE If you use batteries it is forbidden to charge the dry batteries to avoid danger To keep the performance of the instrument and prolong the service life please pay attention to the following 1 The instrument should be avoided to shock hit fall off and other mechanic...

Страница 13: ...he battery out if it is not used for long time to avoid the effect for battery life 7 Must use the charging adapter supported by the model of instrument 7 Faults Solutions Faults Reasons Solutions Cannot turn on the tester Low voltage of battery Charge the battery Big test error Sensor interface is dirty Clean the sensor interface Open light source but no optical power outputs Connector ceramic co...