Troubleshooting
Sentek Pty Ltd
ACN 007 916 672
Page 61
with formatting. The Diviner 2000 data should be backed up before
formatting.
The steps required are describe in section
Formatting the Diviner 2000
page 28.
Ø
To rectify charging problems
Symptom
Remedy
Diviner 2000 display unit fails to
switch on or flashes on momentarily
showing a low battery message
Plug a power pack into the charging
socket and leave on charge for at
least 4 hours. If the Diviner 2000
unit still fails to switch on, unplug
and refit the charger for a second
(and third, if required) 4 hour
charging cycle. If the unit still fails
to charge, see your local
Distributor.
Diviner 2000 display unit fails to
report a full battery status after a 4
hour charging cycle.
Remove the charger plug from the
unit, refit the plug and leave for an
additional 4 hour charging cycle. If
the unit does not report a full
battery status after three 4 hour
charging cycles, it is recommended
the unit and the charger be returned
to your local Distributor for service.
Ø
To undo fatal key errors
A fatal key error may occur if the Diviner unit is not powered off correctly.
1.
Turn on the Diviner 2000.
2.
At the start up screen, press #9990.
3.
This will perform a reset/reboot of the display unit. No data will be
lost
Ø
Diagnosing scan failures
A failed scan can be caused by a number of things.
When a scan fails, verify that the soil profile is correct and ensure that the
full depth of the profile is reached during the scan. Read
To scan (swipe
and go)
on page 19.