Part 1—The Diviner 2000 display unit and probe
Calibrating Diviner 2000
Sentek Pty Ltd
ACN 007 916 672
Page 17
feature…
means…
During installation of the access tube the top
cap lip is generally flush with the soil surface.
Alternatively the lip can be positioned 5 cm/2
inch above the soil surface. This allows the
Diviner 2000 sample to be taken at 5 cm/2
inch below the soil surface rather than 10
cm/3.9 inch. This can be useful for shallow
rooted crops.
The Diviner 2000 Datum value should match
the alignment of the access tube top cap
Datum plate.
Set Diviner Datum value to:
•
0.0 cm when Access tube top cap is
flush with soil surface.
•
5.0 cm/2 inch when Access tube top cap
is 5 cm/2 inch above soil surface.
Soil type
Diviner 2000
allows you to record up to 99
different soil types. Your site details may be
entered using the storage locations ‘01’ to
‘99’. The ‘#’ symbol is used for default
options.
Constant
‘A’, ‘B’ and ‘C’ values
Ø
To create a soil profile (See page 28 to delete a profile)
1.
Use the arrow keys to highlight
Profile
.
2.
Use
NEXT
and
PREVIOUS
to navigate through the different site
profiles or, use the
numeric key pad
to type in the site profile
identification number.
3.
Press
ENTER
to go directly to a particular site profile.
4.
Use the down arrow key to highlight
Datum.
If the top cap is flush with the soil surface, leave the Datum value at
0.0.
If the top cap datum plate is 5 cm/ 2inch above the soil surface, set the
Datum value to 5.0 cm /2.0 inch using the
numeric key pad
and
pressing 5 (or 2).