• Press
or
as many times as required to select the
desired language. Note that changes do not become
effective.
• Press
to confirm the selected language and to deactivate
‘editing mode’. To cancel changes and deactivate ‘editing
mode’ press
or
. Note that upon deactivation of the
‘editing mode’ the ‘Enter’ symbol reappears at the end of the
line and the color of the menu bar changes from yellow to
blue.
Note: After language confirmation, the SDM automatically
exits the menu.
Note: Operator access to the parameter ‘Language’ can
be disabled by the institution by using V-STATS™ within a
password protected area.
Example 3: ‘Confirmation of Membrane Change’
To reset the membrane timer after a successful membrane
change, the membrane change must be confirmed on the SDM
by using the function ‘Membrane Change Done’ in the menu
‘Membrane Change’. To confirm a membrane change, proceed
as follows:
• Press
to access the menu.
• Press
twice to scroll down the blue menu bar to the menu
‘Membrane Change’.
• Press
to open/activate the menu ‘Membrane Change’.
Note: The SDM automatically activates the menu ‘Membrane
Change’ if a sensor with an expired membrane is inside the
Docking Station.
• Press
once to scroll down the blue menu bar to the
function ‘Membrane Change Done’.
• Press
to confirm the membrane change.
Note: The menu ‘Membrane Change’ and the function
‘Membrane Change Done’ are dimmed grey (not accessible)
if the sensor is attached to the patient or inside the Docking
Station. In this case, remove the sensor from the patient or the
Docking Station to confirm the membrane change.