Page 37
.
Patient Monitoring with the SDMS
Note: To terminate the ‘Site time elapsed’ alarm, remove the
sensor from the patient and either press the Enter Button while
the message ‘Sensor off patient (
8
)’ displays or insert the
sensor into the Docking Station.
CAUTION: Do not reattach the sensor to the same site if
any skin irritations are noted during site inspection.
Good to know!
If the safety feature SITE PROTECTION is ON, the SDM
will reduce the sensor temperature to safe values once the
sensor application duration overruns the selected ‘Site Time’
by more than 10% or 30 minutes. PCO
2
/PO
2
are marked as
‘invalid’ thereafter (values replaced by ‘---’).
During monitoring, the current setting for SITE PROTECTION
is indicated in the ‘Sensor Temperature’ icon (p.
48
). A
‘red-blue rightward arrow with tip down’ displays if SITE
PROTECTION is ON, a ‘red rightward arrow’ if it is OFF.
By using V-CareNeT™, for example, remote monitoring and
secondary alarm surveillance of multiple SDMs connected to
the same network as the PC is possible. ‘Operator Events’,
‘Baselines’, and certain SDM settings can be set/controlled
remotely on the included SDMs. Furthermore, download of
SDM Trend Data is simultaneously possible for multiple SDMs.
‘Remaining Monitoring Time’/‘Site Time Elapsed’
Alarm
During monitoring, the ‘Remaining Monitoring Time’ Icon
(p.
48
) continuously indicates the ‘Remaining Monitoring
Time’, i.e. the time until either the selected ‘Site Time’ or – if
PCO
2
is enabled – the ‘Calibration Interval’ elapse (whichever
will occur first).
When the ‘Calibration Interval’ elapses before the selected
‘Site Time’, the ‘Remaining Monitoring Time’ Icon is highlighted
yellow, the message ‘Sensor calibration recommended’ is dis-
played and monitoring is possible another 4 to 6 hours with
PCO
2
marked as ‘questionable’. Thereafter, sensor calibration
is mandatory and PCO
2
and PO
2
are marked as ‘invalid’ (values
replaced by ‘---’). When the ‘Site Time’ elapses, the icon is
highlighted in red and the low priority alarm ‘Site time elapsed’
is triggered. In this case, the sensor must be removed from the
patient for site inspection.