Page 23
.
Patient Monitoring with the SDMS
SP OFF
(or ON and T≤ 41.0 °C in
adults/T ≤ 40.0 °C in neonates)
SP ON
(if T > 41.0 °C in adults/
T > 40.0 °C in neonates)
IH OFF
(or *)
IH ON
(if **)
Note: In ‘Institutional Mode’ it is possible – by using
V-STATS™ – to store up to 4 SDM Profiles on the SDM and select
one of these profiles as ‘Standard Profile’. During subsequent
use, the operator can restore the selected ‘Standard Profile’ (if
modified) or select a different ‘Standard Profile’ in the menu
‘Profiles’. Furthermore, if at power-up of the SDM the LAST
settings differ from those of the selected ‘Standard Profile’,
this menu activates and offers the option to keep the modified
settings, to restore the selected ‘Standard Profile’ or to select
another ‘Standard Profile’.
Good to know!
Various profiles preconfigured by SenTec and tailored to
optimally fit the specific needs of varying clinical settings are
available within V-STATS™.
5
Sensor Temperature: Displays the currently selected
sensor temperature (this indicator is only displayed if the
connected sensor is heated).
WARNING: The use of temperatures higher than 41 °C
requires special attention to patients with susceptible skin,
e.g. neonates, geriatric patients, burn victims, patients with
skin diseases.
6
Special Temperature Settings: Split-arrow indicating
the current configuration of INITIAL HEATING (IH, left part of
arrow) and SITE PROTECTION (SP, right part of arrow).
* ON and T = 44.5 °C in adults/T ≥ 43.5 °C in neonates
** T < 44.5 °C in adults/< 43.5 °C in neonates
7
V-Check™ Mode Indicator: If the V-Check™ Mode
(p.
35
) is ON the ‘V-Check™ Mode Indicator’ displays on the
left of the ‘Sensor Temperature Indicator’
5
and the ‘Special
Temperature Settings Indicator’
6
.
The following information is displayed in the center of the
screen:
Enabled Parameters: Indicates the parameters that are
currently enabled. Ensure to select an option that is approved
for the patient’s age and the intended measurement site
(p.
20
,
21
).
Note: The selectable options depend on the sensor type, the
SDM’s PO
2
activation status, and the selected patient type.