9
4.2
Electrode Input Connections
CAUTION:
• As electrical noise may interfere with electrode signal, please do not route the electrode cable in a
conduit containing AC power wiring.
Electrode Input
Transmitter Terminals
Drive
Red
8
Drive
Ground Black
7
Ground
Temp Green
6
Temp
Temp White
5
Shield
4.3
OC Output
4.3.1
Open Collector Output Connections
Connection to a PLC with built-in power supply
PLC
Power
+
Power Supply-
-
Transmitter Terminals
OC input+
+
4
O.C.+
OC input-
-
3
O.C.-
Connection to a PLC, separate supply
Power Supply
24VDC+
+
24VDC-
-
PLC
Transmitter
Terminals
OC input+
+
4
O.C.+
OC input-
-
3
O.C.-
100K
100K
Electrode Input
Drive
black
Ground
white
Temp
m
p
gr
een
Temp
m
p
red
wire colors in table to left
apply for all
sensors except CS8000TC
and CS615TC, CS620TC.
Wiring For CS615TC,
CS615TC
S855 cable
us
ed with
CS8000TC
Transmitter Terminals
8
Drive
7
Ground
6
Temp
5
Shield
Drive
Ground
Coax braid
Tepm
mp
gr
een
Tem
mp
p
red
Transmitter Terminals
8
Drive
7
Ground
6
Temp
5
Shield
coax center