background image

10 

4.3.2

 

Open Collector Output Operation 

The open collector output can be used as a switch 

or a warning that responds when the process value 

moves above or below a set point, or it can be used 

to generate a pulsing signal with a rate proportional 

to the process value. The output can be disabled if 

not used (select “OFF” in the OC OUTPUT menu). 

The “parameter” mentioned below can be either 

conductivity or temperature. 

Low Mode:

In this mode, the OC output is only active when

the parameter is lower than a user set point. The

output will be inactive when the parameter value is

greater than the set point plus the hysteresis

value.

High Mode:

In this mode, the OC output is only active when

the parameter value is higher than a user

 

set point.

The output will be inactive when the parameter

value is smaller than the set point minus the

hysteresis value.

Proportional Pulsing

In this mode, the OC output will generate a pulse sequence

at the rate defined by the setting in the OC OUTPUT menu.

In the following example, the starting point is 

10

μ

S/cm

, the end point is 

110

μ

S/cm

, and the maximum

frequency is 200 pulses/min:

• The output will be 0 pulses/min when the conductivity value is less than 10

μ

S/cm.

• The output will be 100 pulses/min when the conductivity value is 60

μ

S/cm.

• The output will be 200 pulses/min when the conductivity value is greater than 110

μ

S/cm.

5

 

Menu 

The menu consists of a view menu and an editable menu. The menu has several levels, with the 
view menu being at the topmost level. You can loop within the same level menu items by 
pressing the 

UP

 or 

DOWN

 arrow keys, move to a lower level menu by pressing the 

ENTER 

key, 

and move to an upper level menu by pressing the 

ESC

 key. At any time, the system will return to 

the view menu (default display) if no key is pressed for 10 minutes. 
During normal operation, the view menu is displayed. Use the 

UP

 or 

DOWN

 arrow keys to select 

the information you want displayed. The items will scroll in a continuous loop. System operations 
will not be interrupted during menu interaction. 

OC

OC

OC

Low Setpoint

Hysteresis

Time

Value

N.O.

N.O.

OC

N.O.

:OC Output Active 

:OC Output Inactive  

OC

OC

OC

High Setpoint

Hysteresis

Time

Value

N.O.

N.O.

OC

N.O.

:OC Output Active 

:OC Output Inactive  

Содержание CX105

Страница 1: ...CX105 Conductivity Resistivity Transmitter User Manual REV A 6 Sensorex Corporation USA 11751 Markon Drive Garden Grove CA 92841 U S A www sensorex com...

Страница 2: ...r to ensure against any unintended operation For example an unsafe condition is likely when 1 The transmitter appears visibly damaged 2 The transmitter fails to operate properly or provide the intende...

Страница 3: ...tput Connections 9 4 3 2 Open Collector Output Operation 10 5 Menu 10 5 1 View Menu 11 5 2 Editable Menu 11 5 2 1 Editing Procedure 11 5 2 2 Main Menu 11 5 2 3 Calibration Menu 12 5 2 4 Output Menu 12...

Страница 4: ...d maintain the transmitter 2 Dimension Drawings 2 1 Front and Side View Front view Side view Panel Mount Field Mount 2 2 Installation The CX105 conductivity resistivity transmitters are supplied with...

Страница 5: ...the back of the instrument Make sure that the quick clips are securely attached to the latches 6 Inspect the instrument to make sure that the instrument and the gasket are secured to the panel appropr...

Страница 6: ...possible to install the transmitter on a variety of surfaces 1 Place the gasket on the instrument 2 Thread electrical cables through the connectors on the wall mounting rear cover 3 Connect the power...

Страница 7: ...ut Isolated 4 20 mA output with 0 004 mA 12 bit resolution Update Rate 1 second Maximum Loop Impedance 250 24V 600 31V 1500 48V Memory Non volatile All user settings are retained indefinitely without...

Страница 8: ...built in power supply PLC Terminals Transmitter Terminals 24VDC 2 V Loop 24VDC 4 20mA Loop Input 1 V Loop 4 20mA Loop Input Connection to a PLC with separate power supply Option A Power Supply Transmi...

Страница 9: ...PLC Power Supply Power Supply Transmitter Terminals OC input 4 O C OC input 3 O C Connection to a PLC separate supply Power Supply 24VDC 24VDC PLC Transmitter Terminals OC input 4 O C OC input 3 O C 1...

Страница 10: ...the starting point is 10 S cm the end point is 110 S cm and the maximum frequency is 200 pulses min The output will be 0 pulses min when the conductivity value is less than 10 S cm The output will be...

Страница 11: ...red by pressing the RIGHT DOWN UP DOWN keys in sequence The system will return to the view menu if no key is pressed for 10 minutes Step 2 Navigating the menu with the UP and DOWN keys Once the select...

Страница 12: ...et Temperature 20 0 Sets the value of temperature CONDUCTIVITY S cm Set conductivity 9463 95 S cm Sets the value of conductivity PPM FACTOR 2 00 Set PPM Factor 2 00 Sets the PPM Factor It can be any v...

Страница 13: ...HYSTERESIS 0020 00 S cm As described in 4 3 2 sets the hysteresis value for low mode The OC output is inactive when the conductivity value is greater than the sum of the minimum point and the hystere...

Страница 14: ...he maximum current output to match the external current measurement Adjustable from 19 00 mA to 21 00 mA TEST 4 20mA OUTPUT Press the UP and DOWN keys to manually select any output current value from...

Страница 15: ...e difference is this may take up to 30 seconds Step 3 Set the temperature to the value reading from a precision thermometer Step 4 Set the conductivity to a value which is either from a solution or re...

Страница 16: ...upport 6 3 Parts per Million PPM Factor When the unit in the calibration menu is PPM the following information is useful The PPM factor is a value between 0 01 and 99 99 and the default value is 2 00...

Страница 17: ...05 Conductivity Resistivity Transmitter AR100 11 Mounting Bracket for Transmitter Panel Mount included with CX105 order when lost or damaged AR100 15 Transmitter Gasket included with CX105 order when...

Отзывы: