13
PULSE
the OC mode and relative parameters.
The next two sections (5.2.4.1 and 5.2.4.2) use conductivity as an example, but all of the
operations are the same for temperature.
5.2.4.1
Set 4~20mA Details
Menu Item
Description
Set 4mA Output>
0
00200.
μ
S
Sets the conductivity value for 4mA current loop output.
Set 20mA Output>
2
00000.
μ
S
Sets the conductivity value for 20mA current loop output.
5.2.4.2
OC Output Details
Display
Description
Menu Item
Next Level Menu
MIN ALARM
MIN SETPOINT>
0100.00
μ
S/cm
As described in 4.3.2, sets the minimum point for
low mode. The OC output is active when the
conductivity value is less than the set value.
HYSTERESIS>
0020.00
μ
S/cm
As described in 4.3.2, sets the hysteresis value for
low mode. The OC output is inactive when the
conductivity value is greater than the sum of the
minimum point and the hysteresis value.
MAX ALARM
MAX SETPOINT>
0100.00
μ
S/cm
As described in 4.3.2, sets the maximum point for
high mode. The OC output is active when the
conductivity value is greater than the set value.
HYSTERESIS>
0020.00
μ
S/cm
As described in 4.3.2, sets the hysteresis for high
mode. OC output is inactive when the conductivity
value is less than the maximum point minus the
hysteresis value.
Note: the hysteresis must be less than the
maximum point.
PULSE
RANGE>
000000.
μ
S/cm
200000
μ
S/cm
As described in 4.3.2, sets the start point and end
point for pulse mode.
PULSE RATE>
100 Pulse/min
As described in 4.3.2, sets the maximum pulse rate
for pulse mode.