View Menu
Display
Description
Default display
:
Cond
μ
S/cm
109.55
28.2
℃
Monitors the conductivity and temperature input from the
electrode.
The unit of conductivity or temperature is set in the
calibration menu.
All displays below are temporary. The system will return to the default display if no keys are
pressed in 10 minutes.
4~20mA Output
4.08
mA
Monitors the 4 to 20 mA loop output.
5.2
Editable Menu
5.2.1
Editing Procedure
Step 1. Press and hold the ENTER key for 3 seconds to enter Main Menu:
If a password is required, enter the correct key code. The key code is entered by
pressing the RIGHT-DOWN-UP-DOWN keys in sequence.
The system will return to the view menu if no key is pressed for 10 minutes.
Step 2. Navigating the menu with the UP and DOWN keys.
Once the selected menu is highlighted, press
ENTER
to edit the menu.
Only the highlighted item can be edited.
No parameters will be saved if the
ESC
key is pressed, and the display will return to
the previous menu.
The system will return to the view menu if no activity occurs for 10 minutes.
Step 3. Press the ENTER key to save the new settings and return to Step 2.
NOTE
:
ESC
can be pressed at any time, and the system will exit the current level without saving.
The edited value is effective immediately after pressing the
ENTER
key
.
Repeat steps 2–3 as needed
.
5.2.2
Main Menu
CALIBRATION >
OUTPUT>