Software Installation Instructions
TTM Remote Site Server
©
SENSITECH INC
REV A
9/6/2005
11
Configuration
A. RfNetCfg.xml Refer to Appendix A
1) Browse to the RfNetCfg.xml file in the installed application folder (default c:\program
files\Sensitech\ TTRF Service.
2) Open RfNetCfg.xml with Notepad.
3) <maxthreads> Verify that the value is 3.
4) <Gateway> Set the “id” equal to that of location number which will be written to the
monitor when it enters the network.
5) Set “GwNameOrIp” equal to that of the IP address (static or DHCP) or network machine
name of the gateway being used in the network.
6) Set “TstConnIntvl” equal to 900.
7) Set “ReConnIntvl” equal to 60.
8) <Repeater> Verify that there is a tag for each repeater being installed.
9) Set “HwId” equal to that of the serial number labeled on the repeater.
10) Set “GwBoundTo” equal to that of the IP address or network machine name of the
gateway being used in the network.
Note:
Verify the value does actually exist as a
gateway.
11) <ForceAquireInterval> Verify value is 1800.
Note:
this has to be less than the depart time
in RFSiteCfg. (Section B in Configuration)
12) <Snmp> Set the “MgrNameOrIP” equal to that of the IP Address or network machine
name of the machine running the SNMP Manager (i.e. HP Open View)
13) Verify the “MgrPort” is set to 162.
Note:
Value can only be 161 or 162
14) Verify that if SNMP is being used that both “NotifyLnkDwn” and “NotifyLnkUp” are set to
true.
15) Leave the defaults for “enterprise”, “GwDwnOid”, “DwUpOid”, “RptrDwnOid”, “RptrUpOid”
values.
16) Save all changes to file and close.
B. RfSiteCfg.xml Refer to Appendix B
1) Browse to the RfSiteCfg.xm file in the installed application folder(default c:\program files\
Sensitech\ TTRF Service.
2) Open RfSiteCfg.xml with Notepad.
3) <RFDownloadPath> Set the path to where TTM Desktop is ins \RFDrop
i.e. default would be C:\Program Files\Sensitech\TempTale Manager Desktop\RFDrop
4) <RfApp> Verify there is a tag for each type of application being used in the network.
5) Set “id” equal to that determined in section 3.
6) Set “descript” equal to that of a description of the application. i.e. Closed Loop, Plant to
Shelf
7) Set “download”, “start”, “stop”, “restart”, “mindeptim”
Note:
for Plant to Shelf applications
the mindeptim should be set accordingly. It is the amount of time where a monitor has
considered to have left the network and gone on a trip so data will be downloaded again.
This is more so used in Closed Loop applications.
8) Verify “radiosleep” is set to false.
9) Save all changes to file and close