26
PEAK MODE
OPTIONAL FEATURE
PURPOSE:
Keeping peak readings on the display for situations where the screen cannot be
observed.
Work display peak readings in the
SENSIT GOLD G3
is an optional feature that is used to
maintain peak readings on the display (in addition to the real-time readings) for each sensor
installed. The most common use case for this is when the instrument is being used in a situation
where the display cannot be immediately observed, such as monitoring over-head or confined
space areas, where the instrument is being raised or lowered away from the operator. The peak
reading for each sensor remains on the screen until cleared so that no readings are missed.
TO ENABLE WORK DISPLAY PEAK READINGS:
1. From the working display, press and release the
B
button.
QUICK MENU will appear near
the top of the display.
2. Press and release the
C
button until
PEAK MODE is shown.
3. Press and release the
B
button to enter
PEAK MODE. The working display will be shown
with an additional peak reading in brackets below each real-time reading.
4. Conduct the investigation or survey following federal, state, municipal, and/or company
procedures. The highest measured value for each gas will be shown in brackets under each
live reading. In the case of Oxygen, the lowest reading will be shown as the peak. In the case
of combustible gas, the scale will be shown in addition to the measured value. PPM, LEL, or
VOL will be shown depending on instrument configuration and gas concentration measured.
5. To reset the peak readings, press and hold the
C
button until
AUTOZERO is displayed
(NOTE: this should be done in a gas free environment).
6. To exit the
PEAK MODE, press and hold the
A
button for 2-3 seconds. The peak readings
will be removed, and the standard working display will be shown.