Sensidyne Gilian Gilibrator STABLFLOW Скачать руководство пользователя страница 41

 

Sensidyne Document No. 360-0213-01 - Rev B 

© 2018 Sensidyne, LP  

 Page 35

 

Gilian Gilibrator

® 

3 Dry Cell Calibrator – Operation Manual 

4.6.  

Help and Information 

 

The  Help  Screen  displays  device  information  and  Sensidyne  Support  Contact 

Information. A Help Screen reference display is outlined in the following table. 

 

Display Image 

Description 

 

Help Screen – Displays Serial Number, Last Calibration, 
Calibration  Due  Date,  Cycle  Count  and  Firmware 
Version  Battery  Health,  Filter  Health,  and  Contact 
information for Sensidyne. 
 

Note: The Help Screen will display the information for 
both the Base and the Flow Cell attached. 

 

Help  Screen  Contact  Information  –  For  assistance  or 
service with the Gilibrator 3 Dry Flow Cell Calibration 
System, contact Sensidyne by Phone at 800-451-9444 
or +1 727-530-3602. Sensidyne customer support may 
also be contacted by E-mail at 

[email protected]

 

 

4.7.  

SmartCal 

 

The Gilibrator 3 includes a DB9 port for use with the SmartCal function of the Gilian 

GilAir Plus pumps, by using the accessory cable P/N 780-0015-05-R (Not Included) 

attached to the pump docking station. It can also be used to record live data from the 

unit by attaching it to a null model cable and using a terminal program like PuTTY. 

 

The DB9 is a standard RS232 port and sends data in the familiar Gilibrator2 format at 

2400baud, 8N1. A header is sent on cell reset/connection and it is followed by flow 

data only after the user presses the “START” button. In AVERAGING mode, data is 

only sent after all the samples have been acquired. In CONTINUOUS mode, the data 

is sent after every sample. 

 

In order to use SmartCal, the flow units need to be set to cc/min in the Gilibrator. The 

CONTINUOUS  mode should also  be  selected.  On  the  GilAir  Plus,  select  Gilibrator 

mode for SmartCal under the setup menu. Plumb the pump to the Gilibrator with your 

sample, plug in the cable, set the pump in the dock position closest to the cables, set 

the flow rate, and start the calibration process. The pump will turn on and the Gilibrator3 

will  begin  cycling.  Allow  several  cycles  for  the  system  to  stabilize  and  press  the 

“START” button on the Gilibrator. 

 

As measurements are sent to the pump, it will display the latest flow measurement on 

the screen. The Gilibrator 3 screen will show the measurement once the puck is moving 

up, so it may look delayed from the display of the pump. Allow the system to run until 

the calibration process is done and press the “STOP” button on the Gilibrator. 

Содержание Gilian Gilibrator STABLFLOW

Страница 1: ...ual This Manual Covers All Gilibrator 3 Kit Models Sensidyne Document No 360 0213 01 Rev B 1000 112TH Circle N Suite 100 St Petersburg FL 33716 USA 800 451 9444 1 727 530 3602 www Sensidyne com info S...

Страница 2: ......

Страница 3: ...or 3 Dry Cell Calibrator Operation Manual Gilibrator 3 Identifiers A LCD Display B Status and Notification LED C Ambient Light Sensor and Charge Indicator LED D SD Card Port E Power Switch F Power Por...

Страница 4: ...d manufactured and calibrated in accordance with standards ISO 9001 2015 ISO IEC 17025 2017 ISO IEC 80079 34 ATEX Directive 2014 34 EU and IECEx where applicable Through ongoing review of our designs...

Страница 5: ...uct s operating specifications or 2 is not properly maintained in accordance with its maintenance manual or specifications or 3 has been repaired or modified by persons other than authorized Sensidyne...

Страница 6: ...brator 3 Dry Cell Calibration Kit Descriptions 8 SECTION THREE Set Up 9 3 1 System Components 9 3 2 Base Unit Components 10 3 3 Flow Cell Components 12 3 4 Preparation 13 3 5 System Set Up 14 3 6 Conn...

Страница 7: ...ed without the prior written permission of Sensidyne LP Trademark Notice Sensidyne the Sensidyne logo Gilian and Gilibrator names and logos are registered trademarks of Sensidyne LP Other trademarks a...

Страница 8: ...STORAGE OF THIS PRODUCT BY THE USER PERSON FIRM ENTITY CORPORATION OR PARTY NOT ADHERING TO THE INSTRUCTIONS AND WARNINGS IN THIS MANUAL OR OTHERWISE PROVIDED BY THE SELLER OR FROM NOT ADHERING TO ALL...

Страница 9: ...r draw liquids of any type into the unit Pump failure faulting or user injury may result Do Not operate this unit with corrosive gasses or gasses that condensate Do Not pressurize the calibrator The G...

Страница 10: ...y pack Do not connect the and terminals with metals such as wire Do not carry or store the battery pack with metal objects such as wire necklace or hairpins If the battery pack is short circuited exce...

Страница 11: ...ications Approvals and Compliances The Gilibrator 3 is EN 61010 1 CE RoHS and EMC compliant The Gilibrator 3 contains an internal battery which has been approved for shipping and transport per UN DOT...

Страница 12: ...tor 3 Dry Cell Calibration System include an easy to read color touch screen LCD display multi screen user interface with drop down menus and screen keyboard input capabilities corrections for standar...

Страница 13: ...ent accuracy determines the overall accuracy of the unit The Gilibrator 3 Dry Cell Calibration System consists of two elements the Dry Flow Cell Assembly and the Control Unit Base The function of the...

Страница 14: ...Single Cell Kit 1 Choose 1 Dry Cell 910 1702 US R Gilibrator 3 Low Flow Dry Cell 910 1703 US R Gilibrator 3 Standard Flow Dry Cell 910 1704 US R Gilibrator 3 High Flow Dry Cell Yes Two Cell Kit 1 Cho...

Страница 15: ...timing system This type of microprocessor used in conjunction with the built in software provides an extremely accurate method for calculating the flow rate parameters The Gilibrator 3 Dry Cell Calibr...

Страница 16: ...Light Sensor is located adjacent to the lower right corner of the LCD Display The LED indicates the status of battery charging The sensor allows the LCD backlight to increase to full brightness when...

Страница 17: ...cle The Inlet Fitting is located on the side of the base unit towards the top The inlet is used for makeup air when calibrating a device in suction mode The inlet fitting can also be used to connect a...

Страница 18: ...it takes the puck to travel between a pair of calibrated detectors White LED lights can be seen through the Viewing Window on Dry Cell These lights will illuminate when connected to the base and the b...

Страница 19: ...arrives assembled and has interchangeable Flow Cells that connect into the base unit Figure 3 5 IMPORTANT Before proceeding you MUST charge the battery to full capacity prior to using the calibrator T...

Страница 20: ...a complete Gilibrator 3 Dry Cell Calibration System may be configured Prior to setup make certain you have properly connected the battery charger to the power port on the backside of the Base Control...

Страница 21: ...ure device i e suction pump Tubing must be selected based on the flow range and fittings The calibration base unit comes standard with a 3 8 inch OD fitting Step down adapters are provided with the ki...

Страница 22: ...he base is powered down or when flow cells are disconnected all air paths are bridged to allow free flow of air from the inlet to the outlet and the puck is allowed to drop to the bottom of the cell O...

Страница 23: ...ell type with flow range Reports Screen Reports Screen displays most recent calibration reports by date allows for preview and export of reports and for deletion of single or all reports Settings Scre...

Страница 24: ...ecting Reference Temperature and Pressure Touch the button on the Screen adjacent to the STP icon The Input Keyboard will appear Note User may change from F to C on Settings Display Page Use keyboard...

Страница 25: ...n an averaging set Choose from one of four pre selected values or select to open a keyboard window and select a value from 3 to 20 Type the number ranging from 3 20 of samples you wish to have average...

Страница 26: ...glish Spanish French Portuguese German Dutch Italian Chinese Polish Turk Selecting Screen Brightness Touch the slider along the graduated marks to manually adjust screen brightness Furthest right is B...

Страница 27: ...me Screen and corresponding records Selecting the Ambient Pressure Units Touch the buttons adjacent the Pressure icon to select from four units of pressure in H2O mmHg kPa or hPa Note The Ambient Pres...

Страница 28: ...Description Step 1 Enter the Hour 0 23 Touch Next button Step 2 Enter the Minute 0 59 Touch Next button Step 3 Enter the Month 0 12 Touch Next button Step 4 Enter the Day 1 31 Touch Next button Step 5...

Страница 29: ...tton to initiate the test This may take up to 5 hours do not unplug or turn off the unit while test is in progress A Please Wait window will appear while the test is in progress The unit will need to...

Страница 30: ...t hose barbs Press the OK button Turn the calibrator upside down Wait for the piston inside of the flow cell to fall completely reach the top of the flow cell while upside down Return the unit to upri...

Страница 31: ...the button with the Firmware Update icon A Firmware Update instruction window will appear Press OK button to begin update Do not turn off unit until the update is complete Set Sleep Timer Touch the bu...

Страница 32: ...ned with the SmartCal format on the GilAir Plus Pump Prepare To Ship Touch the button with the Ship Gilibrator icon A Prepare to ship instruction window will appear Disconnect USB Cables Disconnect Po...

Страница 33: ...splayed in the top left corner of the Home Screen The time and date can be edited on the Settings Display page Home Screen Sampling Mode is displayed on the top right corner of the Home Screen Averagi...

Страница 34: ...ays the number of samples within an averaging set Note The sample count window does not appear in Continuous Mode Home Screen Battery Life is displayed in the bottom left corner of the Home Screen The...

Страница 35: ...he averaging sample count data set Note The Gilibrator may take several cycles to stabilize Home Screen Flow Rate displays the live flow rate measurement from the Flow Cell Reset Sample Set Touch the...

Страница 36: ...ord detail Recently Used Pump Screen This screen will populate with recently used pump information If unpopulated or the flow source name is not present touch the button labeled Skip to continue to in...

Страница 37: ...nter in the unique sample identification Touch the button labeled Next to advance to the Operator ID keyboard Pump Operator ID Keyboard Enter in the operator information Touch the button labeled Done...

Страница 38: ...sv If a post cal record is created and exported to the SD card the files will also appear in the same sub folder The post cal files will be denoted with a _ behind the sample ID ex D REPORTS Test 1 Te...

Страница 39: ...port line highlighted touch the Delete button on the display screen Delete record window appears Touch the YES button to delete the single report Preview a Report With the report line highlighted touc...

Страница 40: ...A Please Wait window will appear during data transfer Do not remove the SD card unitl transfer is complete If the file already exists on the SD Card a window will appear asking Overwrite Existing File...

Страница 41: ...rom the unit by attaching it to a null model cable and using a terminal program like PuTTY The DB9 is a standard RS232 port and sends data in the familiar Gilibrator2 format at 2400baud 8N1 A header i...

Страница 42: ...ollowing procedures to keep the unit in proper working order 1 Disconnect all cables from the Base Control Unit 2 Place black caps on both inlet and outlet fittings of Base Control Unit 3 Place black...

Страница 43: ...uck inside the flow cell due to moisture aspiration it must be returned to Sensidyne for service Direct sunlight on the unit may trigger the internal optical sensors and cause the cell to be reset or...

Страница 44: ...2018 Sensidyne LP Sensidyne Document No 360 0213 01 Rev B Page 38 Gilian Gilibrator 3 Dry Cell Calibrator Operation Manual NOTES...

Страница 45: ...S A 800 451 9444 1 727 530 3602 fax 1 727 539 0550 www Sensidyne com info Sensidyne com Authorized EU Representative Schauenburg Electronic Technologies GmbH Weseler Str 35 45478 M lheim Ruhr Germany...

Страница 46: ...1000 112th Circle N Suite 100 St Petersburg FL 33716 USA 800 451 9444 1 727 530 3602 www Sensidyne com info Sensidyne com...

Отзывы: