Sensidyne Document No. 360-0213-01 - Rev B
© 2018 Sensidyne, LP
Page 35
Gilian Gilibrator
®
3 Dry Cell Calibrator – Operation Manual
4.6.
Help and Information
The Help Screen displays device information and Sensidyne Support Contact
Information. A Help Screen reference display is outlined in the following table.
Display Image
Description
Help Screen – Displays Serial Number, Last Calibration,
Calibration Due Date, Cycle Count and Firmware
Version Battery Health, Filter Health, and Contact
information for Sensidyne.
Note: The Help Screen will display the information for
both the Base and the Flow Cell attached.
Help Screen Contact Information – For assistance or
service with the Gilibrator 3 Dry Flow Cell Calibration
System, contact Sensidyne by Phone at 800-451-9444
or +1 727-530-3602. Sensidyne customer support may
also be contacted by E-mail at
4.7.
SmartCal
The Gilibrator 3 includes a DB9 port for use with the SmartCal function of the Gilian
GilAir Plus pumps, by using the accessory cable P/N 780-0015-05-R (Not Included)
attached to the pump docking station. It can also be used to record live data from the
unit by attaching it to a null model cable and using a terminal program like PuTTY.
The DB9 is a standard RS232 port and sends data in the familiar Gilibrator2 format at
2400baud, 8N1. A header is sent on cell reset/connection and it is followed by flow
data only after the user presses the “START” button. In AVERAGING mode, data is
only sent after all the samples have been acquired. In CONTINUOUS mode, the data
is sent after every sample.
In order to use SmartCal, the flow units need to be set to cc/min in the Gilibrator. The
CONTINUOUS mode should also be selected. On the GilAir Plus, select Gilibrator
mode for SmartCal under the setup menu. Plumb the pump to the Gilibrator with your
sample, plug in the cable, set the pump in the dock position closest to the cables, set
the flow rate, and start the calibration process. The pump will turn on and the Gilibrator3
will begin cycling. Allow several cycles for the system to stabilize and press the
“START” button on the Gilibrator.
As measurements are sent to the pump, it will display the latest flow measurement on
the screen. The Gilibrator 3 screen will show the measurement once the puck is moving
up, so it may look delayed from the display of the pump. Allow the system to run until
the calibration process is done and press the “STOP” button on the Gilibrator.