© 2018 Sensidyne, LP
Sensidyne Document No. 360-0213-01 - Rev B
Page 26
Gilian Gilibrator 3
®
Dry Cell Calibrator – Operation Manual
Display Image
Description
Selecting the Communication Format – Touch the
buttons adjacent the
Communication Format
icon
,
to select from three calibrator units communication
types; Gilibrator® 3, Gilibrator® 2, or Defender.
Note: The communication format must be aligned with
the SmartCal format on the GilAir Plus Pump
Prepare To Ship – Touch the button with the
Ship
Gilibrator
icon
.
A
Prepare to ship
instruction window will appear.
Disconnect USB Cables.
Disconnect Power Source (12 VDC Plug).
Press Ok button.
The unit will self-discharge the battery until it reaches
25%.
Upon completion, an instruction window will appear.
Press OK button.
Turn Gilibrator off.
Package appropriately for transport.
Note: Once discharged, do not plug unit back into
power source. You may select the Cancel button at any
time and plug in your unit to recharge back to full
power.