  If  several  errors  are  detected  at  the  same  time  the  different  error  code  numbers  will  be 

added together into one single error code! 

 

PLEASE  NOTE!

 

The  sensor  accuracy  is  defined  at  continuous  operation  (at 

least 3 weeks after installation) 

 

Maintenance 

The 

aSENSE

 

is  basically  maintenance  free.  An  internal  self-adjusting  calibration 

function takes care of normal long term drift associated to the CO

2

 sensor. To secure 

the  highest  accuracy,  a  time  interval  of  five  years  is  recommended  between  CO

2

 

calibrations, unless some special situations have occurred. A zero calibration can be 
performed  by  use  of  pure  nitrogen  or  air  that  has  passed  through  a  chemical 
absorber  and  a  PC  together  with  the 

UIP  software  version  5 

(or  higher).  The  Zero 

Calibration bag can be used to produce carbon dioxide free air. The software can be 
free downloaded from 

www.senseair.com

The RS232-cable and the zero calibration 

bag can be ordered from SenseAir.  The cable  is to be connected to the UART port 
slide  connector  (see  Fig.  5).  For  change  of  control  parameters  and  re-calibration 
(CO

2

 and temperature) this PC tool has to be used. The check can be done on site 

without interfering with the ventilation system. Re-calibration can be done with a PC 
orby shortening of marked holes for at least 10 seconds. 
 
A  Back-ground  calibration  is  an  easy  way  to  correct  a  sensor

’s  zero  point.  The 

sensor  needs  fresh  air  (380-420  ppm  CO

2

).  Shorten  marked  holes  for  at  least  10 

seconds. 
 

This is for sensors with measuring ranges between 0-3000 ppm and 0-4%: 

When a zero calibration shall be executed a plastic tube with 2.2 mm outer diameter 
and 0.8  mm  inner  diameter  shall be  inserted  in  marked  holes  of  the  sensor. Plastic 
tubing is connected to the tube. The gas flow should  be between 0.3 and 1.0 l/min. 
Shorten  marked  holes  for  at  least  10  seconds.  A  succesful  calibration  is 
acknowledged by 5 flashes of the  yellow LED. If unsuccessful, please  release input 
and waits at least 10 seconds before repeating the procedure again. Make sure that 
the sensor environment is steady and calm! 
 

Gas inlet

 

Figure 2 Part of the PCB with holes for gas inlets marked.

 

Содержание aSENSE

Страница 1: ...well as for duct mounting The unit is designed for connecting to Direct Digital Control DDC The linear output functions are pre programmed as CO2 and temperature transmitters with jumper selected outp...

Страница 2: ...o F CO2 value mA 4 2000 16 Temperature value in o C mA 4 50 16 Temperature value in o F mA 4 16 122 32 32 a b c d Value reading a b a d c c a lowest value of the configuration b highest value of the...

Страница 3: ...erature measurements accuracy is valid only for units configured in voltage outputs mode Table II Configuration jumpers for aSENSE Self diagnostics The system contains complete self diagnostic procedu...

Страница 4: ...of Range Error Accompanies most of other errors Resets automatically after source of error disappearance Try sensor in fresh air Check connections of temperature and relative humidity probe Check deta...

Страница 5: ...e cable is to be connected to the UART port slide connector see Fig 5 For change of control parameters and re calibration CO2 and temperature this PC tool has to be used The check can be done on site...

Страница 6: ...n must be taken in order not to damage the temperature probe in the sampling tube When putting the PCB back in the protective housing the probe must be gently positioned in the sampling tube Figure 3...

Страница 7: ...hall ship them to Buyer s designated return destination freight pre paid 2 Warranty Limitations This warranty does not extend to any unit that has been subject to misuse neglect or accident that has b...

Отзывы: