SenseAir aSENSE Скачать руководство пользователя страница 4

 

 

 

 

 

Document 
UMA 139 
 
 

Rev 

Page 
4 (7)

 

 

Error code and action plan

 

 

Bit #  Error 

code 

Error description 

Suggested action 

N/A 

Fatal Error 

Try to restart sensor by power 
OFF/ON. Contact local distributor 

Reserved 

 

 

Algorithm Error

Indicate wrong EEPROM 
configuration. 

Try to restart sensor by power 
OFF/ON.  
Check detailed settings and 
configuration with UIP software 
version 5 and higher. 
Contact local distributor. 
 

Output Error 

Detected errors during output 
signals calculation and 
generation. 

Check connections and loads of 
outputs. 
Check detailed status of outputs 
with UIP software version 5 and 
higher. 
 

16 

Self-Diagnostic Error

Algorithm timeout 

Check detailed self-diagnostic status 
with UIP software version 5 and 
higher. 
Contact local distributor 
 

32 

Out of Range Error 

Accompanies most of other 
errors.  
 
Resets automatically after 
source of error disappearance. 
 

Try sensor in fresh air. 
Check connections of temperature 
and relative humidity probe.  
Check detailed status of 
measurements with UIP software 
version 5 and higher.  

 
See Note 1! 

64 

Memory Error 

Non-fatal error during memory 
operations. 
 

Check detailed settings and 
configuration with UIP software 
version 5 and higher. 
 

128 

Warm Up state 

Is always set after power up or 
power failure. Resets after 
restart sequence. 
 

If it doesn’t disappear in half a 
minute, check power stability. 

 

Note  1

.    Any  probe  is  out  of  range.  Occurs,  for  instance,  during  over  exposure  of  CO2 sensor,  in 

which  case  the  error  code  will  automatically  reset  when  the  measurement  values  return  to  normal. 
Could  also  indicate  the  need  of  zero  point  calibration.  If  the  CO2

 

readings  are  normal,  and  still  the 

error code remains, the temperature sensor can be defect or the connections to it are broken

.  

 

 

Содержание aSENSE

Страница 1: ...well as for duct mounting The unit is designed for connecting to Direct Digital Control DDC The linear output functions are pre programmed as CO2 and temperature transmitters with jumper selected outp...

Страница 2: ...o F CO2 value mA 4 2000 16 Temperature value in o C mA 4 50 16 Temperature value in o F mA 4 16 122 32 32 a b c d Value reading a b a d c c a lowest value of the configuration b highest value of the...

Страница 3: ...erature measurements accuracy is valid only for units configured in voltage outputs mode Table II Configuration jumpers for aSENSE Self diagnostics The system contains complete self diagnostic procedu...

Страница 4: ...of Range Error Accompanies most of other errors Resets automatically after source of error disappearance Try sensor in fresh air Check connections of temperature and relative humidity probe Check deta...

Страница 5: ...e cable is to be connected to the UART port slide connector see Fig 5 For change of control parameters and re calibration CO2 and temperature this PC tool has to be used The check can be done on site...

Страница 6: ...n must be taken in order not to damage the temperature probe in the sampling tube When putting the PCB back in the protective housing the probe must be gently positioned in the sampling tube Figure 3...

Страница 7: ...hall ship them to Buyer s designated return destination freight pre paid 2 Warranty Limitations This warranty does not extend to any unit that has been subject to misuse neglect or accident that has b...

Отзывы: