background image

ENGLISH - 4/8

 

 TYPES OF ERRORS 

Android APP 

configuration

 

Maximum 

measurement field 

errors

 

Calibration 

error

 

Thermal 

coefficient

 

Linearity error

 

EMI

 

TC Input: J, K, E, T, N

 

0.1%

 

0.01%/°K

 

0.2°C

 

<1% (1)

 

TC Input: R, S

 

0.1%

 

0.01%/°K

 

0.5°C

 

<1% (1)

 

TC Input: B (2)

 

0.1%

 

0.01%/°K

 

1.5°C

 

<1% (1)

 

Voltage output (3)

 

0.3%

 

0.01%/°K

 

0.01%

 

(1)

 

Influence of the resistance of the wires: 0.1 

μ

V/

.

 

(2)

 

Zero output for t < 250 °C.

(3)

 

Values to be added to the errors for the selected input.

 ELECTRICAL CONNECTIONS 

CAUTION

 

The upper power supply limits must not be exceeded, as this could cause serious damage 

to the module. Switch the module off before connecting inputs and outputs.

 

To meet the electromagnetic immunity requirements:

 

use shielded signal cables;

connect the shield to a preferential instrumentation earth system;

separate shielded cables from other cables used for power installations 

(inverters, motors, induction ovens, etc...).

 

Make sure 

that the power supply voltage to the module does not exceed: 40 Vdc or 28 Vac,

otherwise the module will be damaged.

 

 USB INTERFACE 

The module has a microUSB connector on the front panel and can be configured using applications 

and/or software programs.

 

For more information, visit www.seneca.it/prod

uc

t

s

/z109tc2-1

 

CU-A-MICRO-OTG 

CU-A-MICROB

 

PC software 

configuration

 

CU-A-MICROB

 

PC

 

t

Содержание Z109TC2-1

Страница 1: ...ponds to the described products and technologies Stated data may be modified or supplemented for technical and or sales purposes INSTALLATION MANUAL Z109TC2 1 Galvanic separation thermocouple converte...

Страница 2: ...NECTIONS Removable 3 way screw terminals 5 mm pitch for cable up to 2 5 mm2 front microUSB POWER SUPPLY Voltage 10 40 Vdc or 19 28 Vac 50 60 Hz Power supply Class 2 Absorption 1 6 W 24 Vdc with 20 mA...

Страница 3: ...prodotti Z204 1 The module must be repaired and damaged parts replaced by the Manufacturer The product is sensitive to electrostatic discharges Take appropriate measures during any operation Important...

Страница 4: ...ous damage to the module Switch the module off before connecting inputs and outputs To meet the electromagnetic immunity requirements use shielded signal cables connect the shield to a preferential in...

Страница 5: ...ltage Applied current 1 External power supply current 2 V output mA output mA output 1 Already powered active output to be connected to passive inputs 2 Non powered passive output to be connected to a...

Страница 6: ...d by the yellow LED t SW1 SELECT THE THERMOCOUPLE TYPE POSITION INPUT POSITION INPUT 1 2 3 4 TYPE 1 2 3 4 TYPE Tc J Tc T Tc K Tc B Tc R Tc E Tc S Tc N KEY ON OFF Tc J Tc K Tc R Tc S Tc T Tc B 2 Tc E T...

Страница 7: ...generator 5 Press START for at least 3 seconds The green LED at the front of the device will flash to indicate that the value has been saved 6 Repeat 4 and 5 for the desired END value 7 Disconnect th...

Страница 8: ...Status LED meaning PWR FAIL Green On The device is powered correctly Flashing 1 flash sec Out of scale Burn Out or internal fault Flashing 2 flashes sec DIP switch setup error ACCESSORIES Code Descrip...