PS-IN202SS Indicator
- 7 -
V2015.05
2.6 Setup Menu Descriptions
This section provides a more detailed description of the selections found in the Setup Menu
Chart. Factory-
set defaults are shown in bold with a checkmark (√).
NAME/CODE
DESCRIPTION
CODE/VALUE
P1
Capacity
Full capacity of the scale. Value should be consistent with legal
requirements and environmental limits on the useful system resolution.
6.0000
√
P2
Accuracy
Accuracy of the scale. Value should be consistent with legal requirements
and environmental limits on the useful system resolution.
Dynamic
P3
Span
Span Gain is related to the A/D integration time. The larger the span, the
higher the internal resolution, but the slower the update speed. Note that
the scale must be recalibrated whenever this parameter is altered. See
Appendix C for more information.
10
√
80
P4
Zero Track
Band
Selects the range within which the scale will automatically zero. Note that
the scale must be in standstill to automatically zero. Selections are in
Display Divisions.
0d
0.5d√
1d, 3d, 5d
P5
Zero Range
Selects the range within which the scale may be zeroed. Note that the
indicator must be in standstill to zero the scale.
100%√
1.9%, 2%, 20%
P6
Power on
Zero Range
Set the zeroing range after the indicator powers on.
0 means deactivate Power On Zeroing
0, 2%, 3%,
20%
√
100%
P7
Motion Band
Sets the level at which motion is detected by comparing the present
display update with the previous one. If motion is not detected for two
seconds or more, the scale is in standstill and can process a Print or Zero
command. Maximum value varies depending on the local regulations.
1d√
3d,
5d
10d
P8
Digital Filter
Averages weight readings to produce higher stability. The higher the filter
setting the greater the stability but the slower the
indicator’s response
time. Choose Med unless a very fast response is needed
FAST
MED
√
SLOW
P9
Overload
Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in P8
“FS” = Full Scale in primary units
FS
FS+ 2%√
FS + 1d
FS + 9d
P10
Calibration
Unit
Selects the primary base unit to be used in the calibration process. Also
the default unit for normal operation.
“1” = calibration unit is lb “2” = calibration unit is kg
1√
2
P11
Zero
Calibration
Places the indicator into the zero calibration routine. Scrolling down with
the
[ZERO]
key one level begins the procedure
Press
[ZERO]
key to
begin sequence
P12
Span
Calibration
Places the indicator into the span calibration routine. Scrolling down with
the
[ZERO]
key one level begins the procedure
Press
[ZERO]
key to
begin sequence
P13
View
Calibration
Actuates the function that allows the user to view both the zero and span
calibration value. The values displayed in this function are valid only after
calibration (P11 and P12) have been successfully completed. Scrolling
down with the
[ZERO]
key one level begins the procedure
Press
[ZERO]
key to
begin sequence
P14
Key-in Zero
Allows the user to key in a known zero calibration value in case of memory
loss in the field. Scrolling down with the
[ZERO]
key one level begins the
procedure
Press
[ZERO]
key to
begin sequence
P15
This sub menu will reset all parameters in
the “P” and “S” menu to the
default setting.
USE WITH CAUTION!
Press
[ZERO]
key twice
to begin sequence