PS-IN202SS Indicator 

- 7 -

V2015.05 

2.6 Setup Menu Descriptions 

This section provides a more detailed description of the selections found in the Setup Menu 
Chart. Factory-

set defaults are shown in bold with a checkmark (√). 

NAME/CODE 

DESCRIPTION 

CODE/VALUE 

P1 
Capacity 

Full capacity of the scale. Value should be consistent with legal 
requirements and environmental limits on the useful system resolution. 

6.0000

 

P2 
Accuracy 

Accuracy of the scale. Value should be consistent with legal requirements 
and environmental limits on the useful system resolution. 

Dynamic 

P3 
Span 

Span Gain is related to the A/D integration time. The larger the span, the 
higher the internal resolution, but the slower the update speed. Note that 
the scale must be recalibrated whenever this parameter is altered. See 
Appendix C for more information. 

10

 

80 

P4 
Zero Track 
Band 

Selects the range within which the scale will automatically zero. Note that 
the scale must be in standstill to automatically zero. Selections are in 
Display Divisions. 

0d 

0.5d√ 

1d, 3d, 5d 

P5 
Zero Range 

Selects the range within which the scale may be zeroed. Note that the 
indicator must be in standstill to zero the scale. 

100%√ 

1.9%, 2%, 20% 

P6 
Power on 
Zero Range 

Set the zeroing range after the indicator powers on. 
0 means deactivate Power On Zeroing 

0, 2%, 3%, 

20% 

√ 

100% 

P7 
Motion Band 

Sets the level at which motion is detected by comparing the present 
display update with the previous one. If motion is not detected for two 
seconds or more, the scale is in standstill and can process a Print or Zero 
command. Maximum value varies depending on the local regulations. 

1d√ 

3d, 
5d 
10d 

P8 
Digital Filter 

Averages weight readings to produce higher stability. The higher the filter 
setting the greater the stability but the slower the 

indicator’s response 

time. Choose Med unless a very fast response is needed 

FAST 

MED

√ 

SLOW 

P9 
Overload 
Limit 

Selects the desired formula which determines the point at which the 
indicator shows overload. All selections are based on the primary unit 
selected in P8 

“FS” = Full Scale in primary units 

FS 

FS+ 2%√ 

FS + 1d 
FS + 9d 

P10 
Calibration 
Unit 

Selects the primary base unit to be used in the calibration process. Also 
the default unit for normal operation. 
“1” = calibration unit is lb               “2” = calibration unit is kg 

1√ 

P11 
Zero 
Calibration 

Places the indicator into the zero calibration routine. Scrolling down with 
the 

[ZERO]

 key one level begins the procedure 

Press 

[ZERO]

 key to 

begin sequence 

P12 
Span 
Calibration 

Places the indicator into the span calibration routine. Scrolling down with 
the 

[ZERO]

 key one level begins the procedure 

Press 

[ZERO]

 key to 

begin sequence 

P13 
View 
Calibration 

Actuates the function that allows the user to view both the zero and span 
calibration value. The values displayed in this function are valid only after 
calibration (P11 and P12) have been successfully completed. Scrolling 
down with the 

[ZERO]

 key one level begins the procedure 

Press 

[ZERO]

 key to 

begin sequence 

P14 
Key-in Zero 

Allows the user to key in a known zero calibration value in case of memory 
loss in the field. Scrolling down with the 

[ZERO]

 key one level begins the 

procedure 

Press 

[ZERO]

 key to 

begin sequence 

P15 

This sub menu will reset all parameters in 

the “P” and “S” menu to the 

default setting. 

USE WITH CAUTION!

 

Press 

[ZERO]

 key twice 

to begin sequence 

Содержание PS-IN202SS

Страница 1: ...Version 2013 08 PS IN202SS Weight Indicator User s Manual...

Страница 2: ...R 2 CONFIGURATION 2 1 SET UP P MENU 2 2 Setup Menu Chart 2 3 Exiting the setup menu 2 4 USER S MENU 2 5 User Menu Chart 2 6 Setup Menu Descriptions 2 7 User Menu Descriptions CHAPTER 3 CALIBRATION 3 1...

Страница 3: ...PS IN202SS Indicator 2 V2015 05 Front View of the Indicator FIG1 shows the front view of the indicator FIG 1...

Страница 4: ...as shown below in FIG 2 Assemble the cables with the pin assignments shown in FIG 2 FIG 2 1 3 CONNECTION TO A PRINTER OR COMPUTER The indicator contains a standard full duplex RS 232 serial port desi...

Страница 5: ...own is the weighing mode c Switch on the indicator The indicator will display P 1 to indicate that it is in Setup P menu mode d Choose the channel you are working on The default channel is 1 You can p...

Страница 6: ...Calibration position Up is the calibration mode Down is the weighing mode c Switch on the indicator The display will go through a digit check then go into Normal Operating mode All front panel keys w...

Страница 7: ...Chart d To move to the selection level press the ZERO right key once The current saved selection is shown e To view the available selections for the current S heading use the TARE down or PRINT up ke...

Страница 8: ...m value varies depending on the local regulations 1d 3d 5d 10d P8 Digital Filter Averages weight readings to produce higher stability The higher the filter setting the greater the stability but the sl...

Страница 9: ...op bit 7E 7 data bits with even parity bit and one stop bit 7n 7 data bits with no parity bit and two stop bits 8n 7O 7E 7n S3 Mode of Serial Transmission Selects when data will be sent out of the ser...

Страница 10: ...int the header information Valid only when S6 is set to 1 0 Do NOT Print Header 1 Print Header 0 1 S12 Back light set Set back light mode OFF ON and Auto when the display is idle the back light will t...

Страница 11: ...nternal A D count and is useful when trying to troubleshoot setup problems b Ensure there are no test weights on the platform and press the ZERO key again to zero out the displayed value c Press the N...

Страница 12: ...libration menu b The display will momentarily show CAL 0 followed by a value This value is the zero calibration value and should be recorded in the table below Press any key to continue c The display...

Страница 13: ...he displayed value to the span calibration value c After setting the exact value press the NET GROSS key to save the value d If the entered value is greater than zero the display will show E CAL 1 mom...

Страница 14: ...isplaying the gross weight T TARE The tare weight has been established in the system lb kg oz PCS lb kg oz PCS Indicates the unit of the displayed weight PCS pieces STABLE This light is illuminated wh...

Страница 15: ...the weight shown on the display 4 5 Tare Function To weigh an item in a container the weight of that container must first be subtracted from the overall weight to obtain an accurate weight reading Thi...

Страница 16: ...ve until all the weight values are added Press and hold ACCU for 2 seconds to enter the accumulation review mode It will show the first value added Press ACCU once and it will move on to the second va...

Страница 17: ...PS IN202SS Indicator 16 V2015 05...