PS-IN202SS Indicator 

- 11 -

V2015.05 

e. If the calibration was successful, the display will show "EndC1" momentarily, then go to
“C2”.

f. Repeat Steps b-e using different test weights then proceed to Step 

“C3”.

g. Repeat Steps b-e using different test weights. Proceed to P12.

h. If the calibration was not successful, one of the error messages below will appear. Take
the indicated action to correct the problem then perform a new calibration.

"

Err0

" - The calibration test weight or the adjusted keyed-in weight is larger than the 

full capacity of the scale. Change the calibration test weight or check the input data. 

"

Err1

" - The calibration test weight or the adjusted keyed-in weight is smaller than 1% 

of the full capacity of the scale. Change the calibration test weight or check the input 
data. 

"

Err2

" - The internal resolution of the scale is not high enough to accept the calibration 

value. 

3.3 View calibration values 
Note: 

The values displayed in this procedure are valid only after a successful calibration has 

been performed using P11 and P12. 
a. While in the Setup mode, scroll to "F 18", then scroll down once using the 

[ZERO]

 key to

enter view calibration menu.

b. The display will momentarily show "CAL 0" followed by a value. This value is the zero
calibration value and should be recorded in the table below. Press any key to continue.

c. The display will momentarily show "CAL 1" followed by another value. This value is the
span calibration value and should also be recorded in the table below. Press any key to
return to upper level (P13).

3.4 Key-in zero calibration value 
Note: 

This procedure is intended for emergency use only in the case of non-volatile memory 

loss. A valid zero calibration value, obtained from a successful P11 calibration procedure, 
must be used. 
a. While in the Setup mode, scroll to "P14" then scroll down once using the 

[ZERO]

 key.

b. The display will momentarily show "CAL 0", followed by a flashing zero. Use the four
direction keys (shown in FIG 1) to adjust the displayed value to the zero calibration value.

c. After setting the exact value, press the 

[NET/GROSS]

 key to save the value.

d. The display will show "E CAL 0" momentarily then revert back to P14.

3.5 Key-in span calibration value 
Note: 

This procedure is intended for emergency use only in the case of non-volatile memory 

loss. A valid span calibration value, obtained from a successful P12 calibration procedure, 
must be used. 
a. While in the Setup mode, scroll to "P15", then scroll down once using the 

[ZERO]

 key.

Содержание PS-IN202SS

Страница 1: ...Version 2013 08 PS IN202SS Weight Indicator User s Manual...

Страница 2: ...R 2 CONFIGURATION 2 1 SET UP P MENU 2 2 Setup Menu Chart 2 3 Exiting the setup menu 2 4 USER S MENU 2 5 User Menu Chart 2 6 Setup Menu Descriptions 2 7 User Menu Descriptions CHAPTER 3 CALIBRATION 3 1...

Страница 3: ...PS IN202SS Indicator 2 V2015 05 Front View of the Indicator FIG1 shows the front view of the indicator FIG 1...

Страница 4: ...as shown below in FIG 2 Assemble the cables with the pin assignments shown in FIG 2 FIG 2 1 3 CONNECTION TO A PRINTER OR COMPUTER The indicator contains a standard full duplex RS 232 serial port desi...

Страница 5: ...own is the weighing mode c Switch on the indicator The indicator will display P 1 to indicate that it is in Setup P menu mode d Choose the channel you are working on The default channel is 1 You can p...

Страница 6: ...Calibration position Up is the calibration mode Down is the weighing mode c Switch on the indicator The display will go through a digit check then go into Normal Operating mode All front panel keys w...

Страница 7: ...Chart d To move to the selection level press the ZERO right key once The current saved selection is shown e To view the available selections for the current S heading use the TARE down or PRINT up ke...

Страница 8: ...m value varies depending on the local regulations 1d 3d 5d 10d P8 Digital Filter Averages weight readings to produce higher stability The higher the filter setting the greater the stability but the sl...

Страница 9: ...op bit 7E 7 data bits with even parity bit and one stop bit 7n 7 data bits with no parity bit and two stop bits 8n 7O 7E 7n S3 Mode of Serial Transmission Selects when data will be sent out of the ser...

Страница 10: ...int the header information Valid only when S6 is set to 1 0 Do NOT Print Header 1 Print Header 0 1 S12 Back light set Set back light mode OFF ON and Auto when the display is idle the back light will t...

Страница 11: ...nternal A D count and is useful when trying to troubleshoot setup problems b Ensure there are no test weights on the platform and press the ZERO key again to zero out the displayed value c Press the N...

Страница 12: ...libration menu b The display will momentarily show CAL 0 followed by a value This value is the zero calibration value and should be recorded in the table below Press any key to continue c The display...

Страница 13: ...he displayed value to the span calibration value c After setting the exact value press the NET GROSS key to save the value d If the entered value is greater than zero the display will show E CAL 1 mom...

Страница 14: ...isplaying the gross weight T TARE The tare weight has been established in the system lb kg oz PCS lb kg oz PCS Indicates the unit of the displayed weight PCS pieces STABLE This light is illuminated wh...

Страница 15: ...the weight shown on the display 4 5 Tare Function To weigh an item in a container the weight of that container must first be subtracted from the overall weight to obtain an accurate weight reading Thi...

Страница 16: ...ve until all the weight values are added Press and hold ACCU for 2 seconds to enter the accumulation review mode It will show the first value added Press ACCU once and it will move on to the second va...

Страница 17: ...PS IN202SS Indicator 16 V2015 05...

Отзывы: