PS-IN202SS Indicator 

- 15 -

V2015.05 

samples on the scale platform then select the sample size and press 

[NET/GROSS]

 to 

confirm.  Sampling is complete.  You can place items on the scale to count.  To exit the 
counting mode, press 

[UNITS] 

to your desired units. 

The scale supports dual counting, which means you can sample on Channel 1 and count on 
Channel 2.  After sampling on Channel 1, press and hold 

[NET/GROSS]

 key to switch to 

Channel 2.  Press 

[UNITS] 

to enter counting mode.  Then you can count on Channel 2.  It 

will keep the sampling results you get from Channel 1.  You can re-sample it anytime on 
either Channel 1 or Channel 2. 

4.7 Accumulation Function 

Please first set the mode in P19 for manual mode or auto mode.  The manual mode is 
default. 

Manual Mode: 

When the readout is stable, press 

[ACCU] 

to add the value in.  The screen will show Add XX 

(XX stands for the sequence number for this value).  If the accumulation function is activated, 
there will be a M+ sign shown on the right upper corner of the screen.  Repeat the above 
until all the weight values are added.   

Press and hold 

[ACCU]

 for 2 seconds to enter the accumulation review mode.  It will show 

the first value added.  Press 

[ACCU]

 once and it will move on to the second value.  Keep 

pressing 

[ACCU] 

until it moves to the ADD UP.  ADD UP is the summary of the total values 

added. 

You may press 

[ZERO]

 any time to return to the weighing mode. 

To clear the memory, press and hold 

[ACCU]

 then press 

[ZERO]

Automatic Mode: 

Zero out the scale.  Place the first weight on the scale.  When the readout is stable, the value 
will be added automatically.  Remove the load and wait for the zero.  Load the second item 
and the value will be added automatically.  Repeat the above until all the weight values are 
added.   

Press and hold 

[ACCU]

 for 2 seconds to enter the accumulation review mode.  It will show 

the first value added.  Press 

[ACCU]

 once and it will move on to the second value.  Keep 

pressing 

[ACCU] 

until it moves to the ADD UP.  ADD UP is the summary of the total values 

added. 

You may press 

[ZERO]

 any time to return to the weighing mode. 

To clear the memory, press and hold 

[ACCU]

 then press 

[ZERO]

Customer Support: 

Toll Free: 844-735-5386

Office Direct: 818-885-5000

 

[email protected]

www.selletonscales.com

Содержание PS-IN202SS

Страница 1: ...Version 2013 08 PS IN202SS Weight Indicator User s Manual...

Страница 2: ...R 2 CONFIGURATION 2 1 SET UP P MENU 2 2 Setup Menu Chart 2 3 Exiting the setup menu 2 4 USER S MENU 2 5 User Menu Chart 2 6 Setup Menu Descriptions 2 7 User Menu Descriptions CHAPTER 3 CALIBRATION 3 1...

Страница 3: ...PS IN202SS Indicator 2 V2015 05 Front View of the Indicator FIG1 shows the front view of the indicator FIG 1...

Страница 4: ...as shown below in FIG 2 Assemble the cables with the pin assignments shown in FIG 2 FIG 2 1 3 CONNECTION TO A PRINTER OR COMPUTER The indicator contains a standard full duplex RS 232 serial port desi...

Страница 5: ...own is the weighing mode c Switch on the indicator The indicator will display P 1 to indicate that it is in Setup P menu mode d Choose the channel you are working on The default channel is 1 You can p...

Страница 6: ...Calibration position Up is the calibration mode Down is the weighing mode c Switch on the indicator The display will go through a digit check then go into Normal Operating mode All front panel keys w...

Страница 7: ...Chart d To move to the selection level press the ZERO right key once The current saved selection is shown e To view the available selections for the current S heading use the TARE down or PRINT up ke...

Страница 8: ...m value varies depending on the local regulations 1d 3d 5d 10d P8 Digital Filter Averages weight readings to produce higher stability The higher the filter setting the greater the stability but the sl...

Страница 9: ...op bit 7E 7 data bits with even parity bit and one stop bit 7n 7 data bits with no parity bit and two stop bits 8n 7O 7E 7n S3 Mode of Serial Transmission Selects when data will be sent out of the ser...

Страница 10: ...int the header information Valid only when S6 is set to 1 0 Do NOT Print Header 1 Print Header 0 1 S12 Back light set Set back light mode OFF ON and Auto when the display is idle the back light will t...

Страница 11: ...nternal A D count and is useful when trying to troubleshoot setup problems b Ensure there are no test weights on the platform and press the ZERO key again to zero out the displayed value c Press the N...

Страница 12: ...libration menu b The display will momentarily show CAL 0 followed by a value This value is the zero calibration value and should be recorded in the table below Press any key to continue c The display...

Страница 13: ...he displayed value to the span calibration value c After setting the exact value press the NET GROSS key to save the value d If the entered value is greater than zero the display will show E CAL 1 mom...

Страница 14: ...isplaying the gross weight T TARE The tare weight has been established in the system lb kg oz PCS lb kg oz PCS Indicates the unit of the displayed weight PCS pieces STABLE This light is illuminated wh...

Страница 15: ...the weight shown on the display 4 5 Tare Function To weigh an item in a container the weight of that container must first be subtracted from the overall weight to obtain an accurate weight reading Thi...

Страница 16: ...ve until all the weight values are added Press and hold ACCU for 2 seconds to enter the accumulation review mode It will show the first value added Press ACCU once and it will move on to the second va...

Страница 17: ...PS IN202SS Indicator 16 V2015 05...

Отзывы: