MANUAL DE INSTRUCCIONES CODIGO 80152 REV E DIC 2019
(Sujetas a modificaciones sin previo aviso) Pag.: 32
J
.P. SELECTA s.a.u.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: www.grupo-selecta.com
Set commands
COMMAND
CONCEPT
MESSAGE
ANSWER
SVT
Set
Temperature (ºC)
#NNSVT<BL>+/-000,0KK<CR><LF>
#NNOKKK<CR><LF> *
SVP
Pressure (mbar)
#NNSVP<BL>0000KK<CR><LF>
#NNOKKK<CR><LF> *
RUN
Run
#NNRUNKK<CR><LF>
#NNOKKK<CR><LF> *
RUNT
Run temperature
#NNRUNTKK<CR><LF>
#NNOKKK<CR><LF> *
RUNP
Run pressure
#NNRUNPKK<CR><LF>
#NNOKKK<CR><LF> *
STOP
Stop
#NNSTOPKK<CR><LF>
#NNOKKK<CR><LF> *
STPT
Stop temperature
#NNSTPTKK<CR><LF>
#NNOKKK<CR><LF> *
STPP
Stop pressure
#NNSTPPKK<CR><LF>
#NNOKKK<CR><LF> *
TBY
Standby time
#NNTBY<BL>00KK<CR><LF>
#NNOKKK<CR><LF> *
TRU
Run time
#NNTRU<BL>00,0KK<CR><LF>
#NNOKKK<CR><LF> *
SRLK
On pump
#NNSRLKKK<CR><LF>
#NNOKKK<CR><LF> *
RRLK
Off pump
#NNTRUKKK<CR><LF>
#NNOKKK<CR><LF> *
INT
Inhibit key pad
#NNINT<BL>0KK<CR><LF>
#NNOKKK<CR><LF> *
INTT
Inhibit temperature key
#NNINTT<BL>0KK<CR><LF>
#NNOKKK<CR><LF> *
INTP
Inhibit pressure key
#NNINTP<BL>0KK<CR><LF>
#NNOKKK<CR><LF> *