INSTRUCTION MANUAL CODE 80152 REV E DEC 2019 (They can be modified without notice)
Page: 31
J
.P. SELECTA s.a.u.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: www.grupo-selecta.com
TBR?
Gets standby set time:
Command
:
#01TBR?KK<CR><LF>
Answer
:
#0100,0KK<CR><LF>
Vaciotem T / Vaciotem TV
TRU?
Gets run time:
Command
:
#01TRU?KK<CR><LF>
Answer
:
#010,00KK<CR><LF>
Answer
:
#0100,0KK<CR><LF>
Time until 9h59mim (9,59)
Time from 10h until 9.99 (99,9)
TRU
Sets run time:
Command
:
#01TRU<BL>0,00KK<CR><LF>
Answer
:
#01TRU<BL>00,0KK<CR><LF>
Answer
:
#01OKKK<CR><LF>
Time until 9h59mim (9,59)
Time from 10h until 9.99 (99,9)
Vaciotem TV
INT
Inhibits/enable key pad (temperature/pressure):
Command
:
#01INT<BL>1KK<CR><LF>
Answer
:
#01OKKK<CR><LF>
1 = inhibit
0 = enable
Vaciotem T/Vaciotem TV
ITT
Inhibits/enable key pad (temperature/pressure):
Command
:
#01INT<BL>1KK<CR><LF>
Answer
:
#01OKKK<CR><LF>
1 = inhibit
0 = enable
Vaciotem TV
ITP
Inibits/enable key pad (temperature/pressure):
Command
:
#01INT<BL>1KK<CR><LF>
Answer
:
#01OKKK<CR><LF>
1 = inhibit
0 = enable
Vaciotem TV
SRLK
Opens pump valve :
Command
:
#01SRLKKK<CR><LF>
Answer
:
#01OKKK<CR><LF>
RRLK
Closes pump valve :
Command
:
#01RRLKKK<CR><LF>
Answer
:
#01OKKK<CR><LF>