  INSTRUCTION MANUAL CODE  80152  REV E       DEC 2019             (They can be modified without notice) 

Page: 1

J

.P.  SELECTA s.a.u. 

 

Ctra. NII Km 585.1 Abrera  08630  (Barcelona) España 

Tel 34 937 700 877   Fax 34  937 702 362

 e-mail: [email protected] -  website: www.grupo-selecta.com

 

  

   VACIOTEM-T  

4001489

   VACIOTEM-TV 

 

4001490

ESTUFA PARA DESECACIÓN AL VACÍO

VACUUM DRYING OVEN

Содержание 4001489

Страница 1: ...out notice Page 1 J P SELECTA s a u Ctra NII Km 585 1 Abrera 08630 Barcelona Espa a Tel 34 937 700 877 Fax 34 937 702 362 e mail selecta jpselecta es website www grupo selecta com VACIOTEM T 4001489 V...

Страница 2: ...com NDICE INFORMACI N GENERAL 3 LISTA DE EMBALAJE 3 ACCESORIOS 3 DESCRIPCI N DEL EQUIPO 4 ESPECIFICACI N T CNICA 4 INSTALACI N 5 OPERACI N 6 TERMOSTATO DE SEGURIDAD 11 COMUNICACI N RS 232 11 RECAMBIO...

Страница 3: ...r duda puede ser aclarada contactando con el servicio t cnico de J P SELECTA s a u 5 ATENCI N NO SE ADMITIR NINGUNA M QUINA PARA REPARAR QUE NO EST DEBIDAMENTE LIMPIA Y DES INFECTADA 6 Toda modificaci...

Страница 4: ...a desecaci n al vac o VACIOTEM T y VACIOTEM TV est n dise adas para permitir la desecaci n para temperaturas regulables desde 35 C hasta 200 C con una estabilidad de 1 C y una homogeneidad de 2 C La r...

Страница 5: ...a si lo hubiera Si el producto a utilizar con la estufa de vac o puede desprender vapores s lo se podr n utilizar conectadas directamente con la estufa bombas de chorro de agua o bombas neum ticas hid...

Страница 6: ...ncionamiento bomba 9 Display lector de vac o 10 Pulsador selector vac o 11 Pulsador selector electrov lvula final de ciclo 12 Pulsador aumentar valor par metros 13 Pulsador disminuir valor par metros...

Страница 7: ...d indicador estado tiempo de espera 19 Display indicador tiempo temperatura 20 Pulsador selector temperatura 21 Pulsador selector tiempo 22 Pulsador aumentar valor par metros 23 Pulsador diminuir valo...

Страница 8: ...s de vac o El display lector del nivel de vac o 9 indica la presi n absoluta presi n atmosf rica en milibares Por lo tanto con la puerta abierta indica la presi n atmosf rica no indica 0 1 Comprobar q...

Страница 9: ...y 9 h 59 min se expresa en unidades sexagesimales A partir de las 10 h se expresa en horas y fracciones decimales de hora cada decimal equivale a 6 min 3 El tiempo de trabajo se empezar a descontar c...

Страница 10: ...resar n d cimas de hora cada d cima representa 6 minutos PUESTA EN MARCHA MODELO T 1 Verificar que se ha seguido correctamente el protocolo de instalaci n 2 Verificar que est cerrada la v lvula de tom...

Страница 11: ...I N RS 232 CARACTER STICAS DE COMUNICACI N RS 232 9600 baudios 7bits de datos 1 bit de stop sin paridad y sin handshake Hasta 15m de cable Cable de conexi n DTE a DTE 9 pins macho ProtocoloASCII con s...

Страница 12: ...ier tienda especializada en compo nentes electr nicos Cable Null modem de 9 pins Los conectores en ambos lados del cable son de 9 pins hembra PC Dispositivo 1 7 8 2 3 3 2 4 6 5 5 6 4 7 1 8 CAMPO DE CO...

Страница 13: ...VT Establece temperatura de consigna SP Set point Comando 01SVT BL 000KK CR LF Respuesta 01OKKK CR LF Vaciotem TV SVP Solicitar presi n de consigna Comando 01SVP KK CR LF Respuesta 010000KK CR LF SVP...

Страница 14: ...P Para el equipo Comando 01STPPKK CR LF Respuesta 01OKKK CR LF Vaciotem T Vaciotem TV ID Solicitar la identificaci n del equipo Versi n software y c digo del equipo Comando 01ID KK CR LF Respuesta 014...

Страница 15: ...00KK CR LF Respuesta 01TRU BL 00 0KK CR LF Respuesta 01OKKK CR LF Tiempo hasta 9h59min 9 59 Tiempo desde 10h hasta 99 9h Vaciotem TV INT Inibir habilitar teclado temperatura presi n Comando 01INT BL...

Страница 16: ...temperatura NNRUNTKK CR LF NNOKKK CR LF RUNP Marcha presi n NNRUNPKK CR LF NNOKKK CR LF STOP Paro NNSTOPKK CR LF NNOKKK CR LF STPT Paro temperatura NNSTPTKK CR LF NNOKKK CR LF STPP Paro presi n NNSTPP...

Страница 17: ...RM3KK CR LF Comandos de lectura COMANDO CONCEPTO MENSAJE RESPUESTA PVT Temperatura PV C NNPVT KK CR LF NN 000KK CR LF PVP Presi n PV mbar NNPVP KK CR LF NN0000KK CR LF SVT Set temperatura C NNPVT KK C...

Страница 18: ...Electrov lvula toma de vac o x1 x1 39005 Resistencia x2 x2 43017 Term stato x1 x1 36027 Rel estado s lido x1 x1 29268 Placa regulador temperatura x1 x0 29269 Placa regulador temperatura vac o x0 x1 15...

Страница 19: ...enquiries please contact your sup plier or J P Selecta s technical service 5 IMPORTANT J P SELECTA WILL NOT ACCEPT ANY EQUIPMENT TO BE REPAIRED IF IT IS NOT DULY CLEANED 6 If any modification eliminat...

Страница 20: ...d start up and temporized run The heating elements surround the working chamber The electronic controller of the VACIOTEM TV have furthermore the vacuum level control and a RS 232 input output The cyl...

Страница 21: ...is on the rear part of the oven 28 Turn the vacuum switch on If the work product can produce vapours you should only use with the oven water jet pumps or hydrorotative pneumatic pumps If an air ballas...

Страница 22: ...Pump indicator led 9 Vacuum level reading display 10 Vacuum selector push button 11 End of cycle electrovalve push button 12 Increasing value push button 13 Reducing value push button 14 Vacuum START...

Страница 23: ...8 Standby time indicator led 19 Time temperature reading display 20 Temperature selecting push button 21 Time selecting push button 22 Increasing value push button 23 Reducing value push button 24 Tem...

Страница 24: ...pressure expressed in mbar HPa So when the oven door is opened it shows the current atmospheric pressure no 0 1 Check that the RUN led indicator 7 does not light Otherwise push the START STOP push bu...

Страница 25: ...placed on the control panel have to stay opened to avoid overpressure inside NOTE The time range between 1 minute and 9 h 59 min is ex pressed in sexagesimal units From 10 hours time is expressed in h...

Страница 26: ...nth part of an hour 6 minutes STARTING UP T MODEL 1 Check that the installation protocol has been correctly followed 2 Check that the ventilation valve 3 is closed 3 Load the material that you have to...

Страница 27: ...hnical service RS 232 COMMUNICATION RS 232 COMMUNICATION OVERVIEW 9600 bauds 7 dat bits 1 bit stop without parity no handshake Up to 15m of cable Cable conection DTE to DTE 9 pins male Protocol ASCII...

Страница 28: ...equal than the transmitted LCR value transmission was good LRC calculation procedure 1 Convert each character on its hexadecimal value 2 Add 3 Take the smallest byte two first characters starting from...

Страница 29: ...KK CR LF Answer 01 000KK CR LF SVT Set point temperature Command 01SVT BL 000KK CR LF Answer 01OKKK CR LF Vaciotem TV SVP Get Set point temperature Command 01SVP KK CR LF Answer 010000KK CR LF SVP Set...

Страница 30: ...mand 01STPTKK CR LF Answer 01OKKK CR LF Vaciotem TV STPP Stops device Command 01STPPKK CR LF Answer 01OKKK CR LF Vaciotem T Vaciotem TV ID Gets device identification Software version and device code C...

Страница 31: ...TRU Sets run time Command 01TRU BL 0 00KK CR LF Answer 01TRU BL 00 0KK CR LF Answer 01OKKK CR LF Time until 9h59mim 9 59 Time from 10h until 9 99 99 9 Vaciotem TV INT Inhibits enable key pad temperatu...

Страница 32: ...RUN Run NNRUNKK CR LF NNOKKK CR LF RUNT Run temperature NNRUNTKK CR LF NNOKKK CR LF RUNP Run pressure NNRUNPKK CR LF NNOKKK CR LF STOP Stop NNSTOPKK CR LF NNOKKK CR LF STPT Stop temperature NNSTPTKK C...

Страница 33: ...00KK CR LF SVT Temperature set C NNPVT KK CR LF NN 000KK CR LF SVP Pressure Set mbar NNSVP KK CR LF NN0000KK CR LF RUN Device status NNRUN KK CR LF NNRUNKK CR LF ID Device identity NNID KK CR LF NN KK...

Страница 34: ...16003 Vacuum gauge x1 x0 16214 Air inlet electrovalve x0 x1 16215 Vacuum electrovalve x1 x1 39005 Heating element x2 x2 43017 Thermostat x1 x1 36027 Rolid Relay x1 x1 29268 Temperature control board x...