MANUAL DE INSTRUCCIONES
CÓDIGO 80381 REV C
12/2020
Pag.: 26
J
.P. SELECTA s.a.u.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail:[email protected] - web site: www.jpselecta.es
STARTING UP
1. Check that the installation protocol has been correctly followed.
2. Switch on the main switch (see picture). The graphic screen of
the temperature controller will light up.
NOTE: Maximum power you should connect on inner sockets is
500 W.
3. Configure the clock (real time clock) only the first time or when the
summer / winter time change occurs. See “Clock configuration”.
Selection offset value
“Offset” option allows increasing or reducing temperature reading in
order to adjust it at other readings made with a standard temperature.
1 Check that the equipment is in RUN mode (in operation)
2 Follow the adjustment instructions to calibrate.
Temperature alarm / electronic safety thermostat
Control alarms can be set to show an alarm message as a warning,
telling that temperature values reached are out of programmed range.
The equipment will continue normally operating despite the error
condition. During the error condition it will be displayed on the screen
and when the temperature is restored within the limits, the “alarm”
status will disappear from the screen.
Note: Be aware when programming the alarm values, that opening the
cabinet door may cause the temperature to reach values outside the
programmed range and then the display will show the alarm message.
USB LOG
The device has an internal memory for data acquisition (Datalogger).
Once the cycle is finished, it is possible to download this informa
-
tion to an USB memory via an USB connector. The device (USB
memory) must be formatted to FAT 32.
Functioning:
• By default, the registrar is not active.
Follow the instructions to change the log interval (see settings>log).
Adjust this interval with the conditions of your program.
With the cycle finished:
• Connect the «USB memory» to the USB connector on the front of
the device.
• When an «USB memory» is connected, the registration data will