OPTICAL AND FIELD MAGNIFCATION TABLE

CLEANING

TROUBLESHOOTING TIPS

The optical magnifications achieved with the Colposcope are determined by 3 variables: the objective focal distance, the 
position of the Magnification Selector and the by 
eyepiece used.

The table shows the optical magnification and the observed field diameters, in millimeters, depending on those 
variables. At the bottom line of the table the illuminated field is shown, corresponding to the light disk incident on the 
field, with a diameter that depends only on the focus of the work objective.

The magnification selector has 6 positions: 1, 2, 3, 4, and 5. The 3 position is repeated in the selector. The working 
position is the one aligned to the red dot mark.

The Objective, pointed to the field under clinical or surgical treatment, is exposed to blood and other bodily fluids. Spots 
on the objective dim the passage of the light.  Stains blur the light, reducing the luminosity of the optical observation.

Cotton padding wetted in alcohol rubbed softly, in CIRCULAR MOVEMENTS over the optical surface, is enough to clean 
the stains, except those of clotted blood. Clotted blood shall be cleaned with cotton and hydrogen peroxide.
 
If the eyepieces are too dirty, change the cotton at each circular friction motion to avoid spreading out the dirt.
 
On the other hand, there is an effective protector for the objective, fitted under pressure at the outer edge of the lens 
which protects the objective against “splatter” and contamination.  
 
To have the eyepieces and objective well cleaned for clinical use, it is sufficient to use cotton.
 
The metallic parts (chrome-plated or painted) can be cleaned with cotton and alcohol.

ELECTRIC CONNECTIONS
 

Ensure that none of the following are present:

 

Broken wires inside the cable.

 

Loose wires at the connector.

 

Broken wires at the fixation, inside the connector.

 

Oxidized connectors terminals.

Magnifica(on

Field

Magnifica(on

Field

Magnifica(on

Field

Magnifica(on

Field

Magnifica(on

Field

1(0.4)

3.89

46.32

3.40

52.94

2.72

66.18

2.27

79.41

1.70

105.88

2(0.6)

5.83

30.88

5.10

35.29

4.08

44.12

3.4

52.94

2.55

70.59

3(1.0)

9.71

18.53

8.50

21.18

6.80

26.47

5.67

31.76

4.25

42.35

4(1.6)

15.54

11.58

13.60

13.24

10.88

16.54

9.07

19.85

6.80

26.47

5(2.5)

24.29

7.41

21.25

8.47

17.00

10.59

14.17

12.71

10.63

16.94

1(0.4)

4.86

41.18

4.25

47.06

3.40

58.82

2.83

70.59

2.13

94.12

2(0.6)

7.29

27.45

6.38

31.37

5.10

39.22

4.25

47.06

3.19

62.75

3(1.0)

12.14

16.47

10.63

18.82

8.50

23.53

7.08

28.24

5.31

37.65

4(1.6)

19.43

10.29

17.00

11.76

13.60

14.71

11.33

17.65

8.50

23.53

5(2.5)

30.36

6.59

26.56

7.53

21.25

9.41

17.71

11.29

13.28

15.06

1(0.4)

6.22

41.18

5.44

47.06

4.35

58.82

3.63

70.59

2.72

94.12

2(0.6)

9.33

27.45

8.16

31.37

6.53

39.22

5.44

47.06

4.08

62.75

3(1.0)

15.54

16.47

13.60

18.82

10.88

23.53

9.07

28.24

6.80

37.65

4(1.6)

24.87

10.29

21.76

11.76

17.41

14.71

14.51

17.65

10.88

23.53

5(2.5)

38.86

6.59

34.00

7.53

27.20

9.41

22.67

11.29

17.00

15.06

Magnifica(on   Table   for   985

Binocular   Head   Focal   Length:   F=170mm

   Eyepiece

Posi(on   of   

BuNon

Objec(ve   Lenses

175

200

250

300

400

90

108

144

10X/18

12.5X/16

16X/16

Illuminated   Field

65

72

Содержание Coolscope 985 LED

Страница 1: ...OWNER S MANUAL PRECISION MICROSCOPES COLPOSCOPE MODEL 985 LED ...

Страница 2: ...e 800 489 2282 Local 314 968 2282 Fax 314 968 3601 E mail micro seilerinst com www seilerinst com Microscope Division Manager Dane Carlson ext 365 dcarlson seilerinst com Corporate Marketing Sales Representative Nicole Rasch ext 345 nrasch seilerinst com Technical Product Information Tony Leise ext 363 tleise seilerinst com Bulb Order Information Sandra Jeremich ext 336 sjeremich seilerinst com PR...

Страница 3: ...SE 4 TECHNICAL DATA 4 BINOCULARS 5 OBJECTIVE 5 COLPOSCOPE ASSEMBLY 6 COLPOSCOPE HANDLING 7 HOW TO FOCUS THE COLPOSCOPE 7 OPTICAL AND FIELD MAGNIFCATION TABLE 8 CLEANING 8 TROUBLESHOOTING TIPS 8 LED LIGHTSOURCE 9 OWNER S RECORD 15 NOTES 15 ...

Страница 4: ... colposcope Do not allow persons unfamiliar with the equipment to operate the colposcope A Colposcope Head Microscope Galilean System Binoculars Straight F 170MM optional Inclined F 135 mm Objective focus 300 mm Eyepieces Wide angle type 12 5x adjustable 6 to 6 dioptries Magnification See table Observation field See table Illumination field See table Interpupillary distance Adjustable between 42 m...

Страница 5: ...dy 2 Eyepieces with diopter adjustment 3 Eye Protectors 1 3 2 OBJECTIVE The Colposcopes are factory shipped with the following objective F 300 mm The following optional objectives are available F 175 mm F 200 mm F 250 mm F 400 mm ...

Страница 6: ... the intake of lateral lights and create a suitable darkroom to the observer Precautions The Binoculars contain optical parts internally that cannot be cleaned without disassembly It is therefore advisable to keep the eyepieces inserted in the eye tubes at all times in order to keep impurities from getting inside the housing It is recommended that if the eyepieces are removed a clean cloth be plac...

Страница 7: ...distance of 300 mm keep a gap of about 300 mm between the lens and the specimen NOTE It is easier to focus when the illumination system is on The illuminated field coincides with the focused object Parfocalizing Procedure 1 Turn the magnification drum to the highest setting 5 and focus sharply on a target such as a piece of paper with an X drawn on it by using the focusing knobs 2 The turn the mag...

Страница 8: ...tive against splatter and contamination To have the eyepieces and objective well cleaned for clinical use it is sufficient to use cotton The metallic parts chrome plated or painted can be cleaned with cotton and alcohol ELECTRIC CONNECTIONS Ensure that none of the following are present Broken wires inside the cable Loose wires at the connector Broken wires at the fixation inside the connector Oxid...

Страница 9: ...n or nitrate oxide To prevent any potential electro magnetic interference do not use any kind of cellular phone near the light source This product should be used only with type BF endoscopic instruments which have been certified according to IEC 601 1 for medical equipment and IEC 601 2 18 for endoscopic equipment Caution This product is not provided as sterile Caution All devices connecting to th...

Страница 10: ...eration Continuous operation Water Resistant Not Protected Equipment IPX0 Operating Environment Temperature Relative Humidity Air Pressure 0 to 40 C 32 to 104 F 0 to 85 rh non condensing 700 to 1060 hPa Storage Environment Temperature Relative Humidity Air Pressure 20 to 60 C 4 to 140 F 30 to 95 rh non condensing 700 to 1060 hPa Dimensions 4 65 W x 3 32 H x 7 25 D Weight 2 5 lbs 1 14 kg 4 OPERATIN...

Страница 11: ...ction 1 AC main inlet Accepts AC power cord 2 Fuse 2A 250V 1 4 x 1 time delay 3 Grounding Connector For potential equalization 4 Venting grids Allows adequate ventilation and cooling of the light source 4 3 Bottom Panel Figure 3 LED Light Source Product Label 2 3 1 4 ...

Страница 12: ...his equipment may be subjected to radio interference To avoid adverse electromagnetic effects DO NOT operate this equipment near RF energy equipment 5 2 CONNECTING THE FIBER OPTIC CABLE Turn intensity control to the minimum setting Connect the light cable to the instrument and then plug the light cable endfitting into the port on the front panel 6 OPERATION 6 1 POWERING UP THE LED LIGHT SOURCE To ...

Страница 13: ...nty from the date of shipment on workmanship and all defects of materi al excluding replaceable sweatbands headband fiberoptic cable and lamps Should your product prove to have such defects within one year of the shipment Seiler Instrument will repair or replace the product or component part without charge Should your LED Light Source product s need servicing under this warranty please contact Sei...

Страница 14: ... be found at http www eiae org Please contact customer service to issue a return authorization to return product to manufacturer at the end of product life 11 TROUBLESHOOTING Problem Solution The power indicator refer to 4 1 is not lit A Check that the AC power cord is properly connected B Check the unit fuses If necessary replace The power indicator is lit but the lamp will not ignite Turn intens...

Страница 15: ...15 Model Number Serial Number Date Purchased Location Purchased OWNER S RECORD NOTES ...

Страница 16: ...PRECISION MICROSCOPES 3433 Tree Court Industrial Blvd St Louis Missouri 63122 Toll Free 800 489 2282 Local 314 968 2282 Email micro seilerinst com Web www seilerinst com Version 09 22 14 ...

Отзывы: