21
1-3 Electrical Specifications
Accuracy is given as ±(% of r counts of least
significant digit) at 23°C ± 5°C, with relative humidity less
than 80% R.H.
ACV and ACA specifications are ac coupled, true R.M.S.
for square wave, accuracy is unspecified.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.) :
Add 1.0% for C.F. 1.0 ~ 2.0.
Add 2.5% for C.F. 2.0 ~ 2.5.
Add 4.0% for C.F. 2.5 ~ 3.0.
Max. Crest Factor of Input Signal :
2.0@4500 counts
1.5@6000 counts
Frequency Response is specified for sine waveform.
Position Error of Clamp:
±1.5% of LCD reading,
unspecified when the conductor is outside 5mm from the
optimum position.
DC Voltage
Range
Resolution
Accuracy
600.0mV
0.1mV
±(
0.7% +5D)
6.000V
0.001V
±(
0.7% +2D)
60.00V
0.01V
600.0V
0.01V
1000V
1V
Input Impedance:
10M
Ω
Overload Protection:
AC/DC 1000V