Figure 9.2
9.6 Use of the template
The comparison with the template is the stop criterion of the data acquisition. This comparison shall
be made with the «
Template Channels
» selected in the
page.
The information window at the bottom of the page
sums up the general configuration of the data acqui-
sition:
•
channels and functions between validated channels for data acquisition
•
channels and functions between channels compared to the template
•
number of points per channel (as a function of the number of blocks)
•
total data acquisition time (as a function of the acquisition speed)
•
position of the trigger (as a function of pre-trigger phase)
Figure 9.3
The comparison to the template is only possible when the channels are validated for data acquisition. An error
message shows up if you have selected « Template Channels » that have note been validated.
The data acquisition is made as above (see chapter
Memory Mode
).
When the memory block is full, the device checks that all measured points on the template channels are between the 2
min. and max. reference curves constituting the template.
If all points are between the curves, the data acquisition restarts (in the same block). If not, the data acquisition stops.
Содержание BK Precision DAS30
Страница 9: ...9 17 Specifications 81 18 LIMITED TWO YEAR WARRANTY 82 19 Service Information 83 ...
Страница 22: ...22 Figure 4 4 Figure 4 5 You get access to ...
Страница 63: ...63 Figure 11 10 ...
Страница 67: ...67 Figure 12 5 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 74: ...74 Figure 14 3 ...