![SEFRAM BK Precision DAS30 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/sefram/bk-precision-das30/bk-precision-das30_user-manual_1233326043.webp)
43
Trigger mode
Double Trigger
; see chapter
Memory mode
.
8.1 Restrictions
Real time data acquisitions to a file are limited by the transfer rate between the inputs and the saving means of the
internal memory or USB sticks.
It is advisable no to create too big files, because reading them may be very long. It takes ca. 3 minutes to read a 200MB
file.
The minimum transfer rate to the flash disk is 100 Kwords/s (10µs)
To know the number of channels you can record at a given acquisition period, calculate:
number of channels = 100 000 x acquisition period example: at 20µs, the number of channels is max. 2.
Содержание BK Precision DAS30
Страница 9: ...9 17 Specifications 81 18 LIMITED TWO YEAR WARRANTY 82 19 Service Information 83 ...
Страница 22: ...22 Figure 4 4 Figure 4 5 You get access to ...
Страница 63: ...63 Figure 11 10 ...
Страница 67: ...67 Figure 12 5 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 74: ...74 Figure 14 3 ...