User Guide
©2008
-2021 Seeed Technology Co., Ltd. All rights reserved. solution.seeedstudio.com
9
/
22
1
)
EC Calibration failed and EC value is incorrect.
2)
Can not communicate with sensor due to forgetting the communication parameters.
5.5 Calibration
When performing the EC calibration, you should guarantee the temperature of the standard EC
calibration solution is around 25°C. And wait for a while for temperature and EC equilibrium after
immersing the sensor into the standard solution.
5.5.1 Perform EC calibration by on board buttons
1
)
Short press SW1-SET button and enter into the EC calibration mode, then the led indicator flash
two times every second, Note that communication is not available in calibration mode.
2
)
In EC calibration mode, Immerse the electrode in 1413us/cm solution for a while and short press
the SW2-CALL button to perform the auto calibration.
3
)
In EC calibration mode, Immerse the electrode in 12880us/cm solution for a while and short
press the button to perform the auto calibration.
4
)
Short press SW1-SET button to exit the EC calibration mode
5
)
Verify the sensor output.
5.5.2 Perform EC calibration by command
1
)
Immerse the electrode in 1413us/cm solution for a while and then write 0xFFFF to modbus
register 0x0030 (ECCALIB_1413 EC calibration point for 1413us/cm) to perform auto calibration.
2
)
Immerse the electrode in 12880us/cm solution for a while and then write 0xFFFF to modbus
register 0x0031 (ECCALIB_12880 EC calibration point for 12880us/cm) to perform auto
calibration.
3
)
Verify the sensor output.
5.5.3 Perform EC calibration by SensorOneSet Configuration Utility
SensorOneSet is a configuration utility to read/set sensor config for all of our serial
communication sensor products. Please contact us if you need the English version.