User Guide
©2008
-2021 Seeed Technology Co., Ltd. All rights reserved. solution.seeedstudio.com
6
/
22
ALL RS485 communication parameters (Modbus Slave Address, baud rate, parity, data bits,
stop bits) are set in internal register and can be saved when power down, the default setting is
ADDRESS=1, BAUDRATE=9600bps, PARITY=NONE, DATABITS= 8bits,
STOPBITS=1bit;
Sometimes you may FORGET the communication settings. In this case, you can open the
shield module and press the SW1-SET button at least 3 seconds, then all the communication
parameters reset to default setting, then communicating with the sensor using the default
setting to set your desired settings.
Please re-power up the sensor to make the settings effective.