5 Installation , On board buttons and Calibration
5.1 Installation
1/2"NPT screw threads installation. Please refer to the dimensions.
5.2 Installation Guide
Should following the common requirements for conductivity electrode installation.
5.3 On board buttons
There are 3 buttons on the transmitter board.
Button
Functionality
Comment
SW1-SET
1
)
Enter into EC calibration
2
)
Exit EC calibration
3
)
Reset to factory settings
1
)
Long press for more than 3
seconds to restore the factory
settings.
2
)
Short press to enter into the EC
calibration.
3
)
Short press to exit the EC
calibration.
SW2-CALL
EC calibration for
1413us/cm(1.413ds/m)
In EC calibration mode, Immerse
the electrode in 1413us/cm solution
for a while and short press the
button to perform the auto
calibration.
SW3-CALH
EC calibration for
12880us/cm(12.88ds/m)
In EC calibration mode, Immerse
the electrode in 12880us/cm
solution for a while and short press
the button to perform the auto
calibration.
5.4 Restore Factory Settings
Long press SW1-SET button for more than 3 seconds to restore the factory settings. You can
restore the factory settings in following cases: