User Guide
©2008
-2021 Seeed Technology Co., Ltd. All rights reserved. solution.seeedstudio.com
14
/
22
Data Range
0-20000 For 0-20000mg/L
Default: N/A
Power Down Save
N/A
Note:
SALINITY
Example:When REGISTER = 0x0702 (HEX format), then VALUE=(0x07*256+0x02)=1794mg/L,
Salinity is derived by EC, SALINITY=EC*
SALINITYCOFF, in which SALINITYCOFF is a
coefficient, please refer to SALINITYCOFF.
TDS---Total Dissolved Solid
Data Range
0-20000 For 0-20000mg/L
Default: N/A
Power Down Save
N/A
Note:Total Dissolved Solid
Example:When REGISTER = 0x0702 (HEX format), then VALUE=(0x07*256+0x02)=1794mg/L,
TDS is derived by EC, TDS=EC*
TDSCOFF, in which TDSCOFF is a coefficient, please refer to
TDSCOFF.
ECRAWAD
Data Range
0-4000
Default: N/A
Power Down Save
N/A
Note: Conductivity raw AD value
Example:When REGISTER = 0x0702 (HEX format), then VALUE=(0x07*256+0x02)=1794
TEMPCOMPENSATEEN
Data Range
0: External Temperature Sensor
1: Onboard temperature sensor
2: Disabled
0
Power Down Save
YES
Note: Temperature compensation
ECTEMPCOFF---EC Temperature Compensation Coefficient
Data Range
0-100 for 0.0%-10.0%
Default: 20(2%)
Power Down Save
YES
Note:EC Temperature Compensation Coefficient
SALINITYCOFF---Salinity Coefficient
Data Range
0-100 for 0.00-1.00
Default: 55(0.55)