21. UNIT LABEL DRAWING
[mm]
P52
Страница 1: ...counter with calculation function DT 601CG www shimpoinst com info shimpoinst com Operation Manual Read this manual thoroughly prior to usage Use this instrument only after reading the manual completely Follow all safety precautions ...
Страница 2: ...y However depending on the situation it might lead to greater outcome The act indicated by this sign is strictly prohibited The act indicated by this sign must be executed by no means BE CAREFUL NOT TO GET ELECTROCUTED DO NOT DROP OR SHAKE KEEP AWAY FROM ELECTRIC WIRE DO NOT TOUCH TERMINALS WHILE THE POWER IS ON Safety Precautions DO NOT DISSEMBLE OR TOUCH INTERNAL PARTS WHILE THE POWER IS ON Belo...
Страница 3: ...electric voltage pulse input and DC power supply Model type will be DT 601CG AI F DC e g2 Choose analog voltage output 0 to 5V 90 phase co ntrast input sensor power supply DC24V Model type will be DT 601CG RE AV4 S24 INDEX PAGE PAGE 1 Summary 2 17 BCD output 48 2 Specification 4 18 BCD input 49 3 Meter installation 6 19 Outline dimensional drawing 50 4 Names and functions of each parts in the fron...
Страница 4: ...ing condition Up Down and Automatic reset setting will be invalid OFFSET 0 OFFSET 0 PRESET B OFFSET PRESET A No automatic reset With automatic reset n Minimum count Display Continuous without re output with re output n Minimum count Display No automatic reset With automatic reset Continuous Continuous Continuous n Minimum count Display without re output with re output Continuous Continuous OFFSET ...
Страница 5: ...rinciple 1 shot or continuous shot will be output from OUT1 to OUT4 under the condition of count display Preset A PresetB However if 2stages preset operation or 1stage preset operation was chosen and used with the condition of with re output or without automatic reset the condition will be as below irrespectively of the mode chosen UP DOWN Output set ON continuous 1 shot Count display Preset A Pre...
Страница 6: ... collector pulse output or actual contact output Output from terminal block 9 6 OUT1 and 10 6 OUT2 Two points NPN open collector pulse output Maximum rating 6 is GND common However with the line receiver it will not function Output from terminal block 15 16 OUT3 and 17 18 OUT4 Two points PhotoMOS relay a contact output Rated load current stabilization output Terminal block 4 3 to be ON more than 5...
Страница 7: ... From terminal block 19 20 Load resistance more than xxx Load resistance les than xxx Within 0 3 23 C against displayed value absolut e value 100ppm C about 50ms as a duration to reach 90 output change 12bit D A conversion system 4000 resolution maximum output range up to 102 4 of each output s maximum value with analog output calculation is done against the displayed value with 7 segments LED Thi...
Страница 8: ... Insert the attachment lugs in both sides of the meter 3 Slide the attachment lugs to the back towards the terminal block side and turn the screw to steady the meter both sides NOTE 1 Make sure to horizontally installed 2 Panel has to be 1 0mm to 4 0mm thick 3 Do not tighten the screw too much The case might break Meter Installation panel Splash proof packing Attachement parts Attachement parts Dr...
Страница 9: ...rning on TEST mode is activated when power is turned on while pressing this key To escape from TEST mode turn off the power While measuring Mode setting is activated when Shift key is pressed more than 2seconds while pressing this key Preset value setting is activated when the key is pressed more than 2seconds Display offset value setting is activated when Up key is pressed more than 2seconds whil...
Страница 10: ... or modify mode protect function 12 Enter key While turning on Format the each set values by turning on while pressing this key While measuring Change the display from Display 1 D1 and Display 2 D2 While setting While setting value will be registered with this key and return to the measuring display 13 Reset key While measuring Switch back the display to ZERO or cancel the alarm output Switch back...
Страница 11: ...ing on the sensors Refer to the wiring diagram on P11 Maximum power supply to sensor is DC12V 100mA optionally DC24V 60mA never over load Wrong wiring might cause damages to sensor or circuit 4 Make sure to tighten the screws on the terminal block 5 About BCD output please refer to BCD output on P48 About BCD input please refer to BCD input on P49 6 Sensor s power source should not be used for oth...
Страница 12: ...C24V 60mA never over load Wrong wiring might cause damages to sensor or circuit 4 Make sure to tighten the screws on the terminal block 5 About BCD output please refer to BCD output on P48 About BCD input please refer to BCD input on P49 6 With line receiver input L1 L2 type terminal block 9 and 10 will be input terminal So alarm output OUT1 and OUT2 cannot be used 7 Sensor s power source should n...
Страница 13: ... wire style pulse sensor B DC two wire style pulse sensor C Contact output sensor D 90 phase contrast input E Sinusoidal Sine curve input F Line receiver input G Three wire current modulated pulse sensor H Two wire current modulated pulse sensor P11 ...
Страница 14: ...RMAT OF INPUT CIRCUIT 1 Sensor input NPN open collector pulse input or voltage contact input 2 Sensor input Electric pulse input 3 Reset External input NPN open collector pulse input or or Sensor Power P12 ...
Страница 15: ...7 FORMAT OF OUTPUT CIRCUIT 1 Alarm input OUT1 OUT2 NPN Open collector pulse output 2 Alarm output OUT3 OUT4 PhotoMOS relay output P13 ...
Страница 16: ...ory preset mode Sinusoidal input V3 sine curve input N and line receiver input L1 L2 3 For 90 phase contrast input RE type make sure to keep the factory preset mode HI for both A B inout for respond frequency 4 Dip switch setting must be done by the combinations shown on the above chart The usage of the combination which is not indicated above might cause some error black shows the setting NPN Ope...
Страница 17: ... times Push the Up key for 2 seconds while pushing the Mode key Setting the Ofset 1 Setting the Ofset 2 Setting the A Out 1 3 Setting the B Out 2 4 Push the Mode key for 2 seconds ENT Record and finish RS T Unrecord and finish Setting the BCD Input Setting the BCD Output Setting the Display of Max Analog output Setting the Analog sDisplay Setting the RST key Setting the External Input Funcion Sett...
Страница 18: ...ing value Initialization Initialization can be done by turning the power on while pressing ENT key After the initialization set value will be as chart 2 3 4 and 5 calculation holding data and batch count data will be set as ZERO Caution Initialization makes every current set value to be reset Record the current value prior to the initialization If the internal computer has interference due to some...
Страница 19: ... B will show 00 and mode 00 will be applied Change the position of flashing digit One hit makes 1digit move right hand side Change the value of flashing digit One hit makes the value bigger by 1 Some digit might not go up til 9 depending on the setting item Change the value of flashing digit One hit makes the value smaller by 1 Some digit might not go up til 9 depending on the setting item Change ...
Страница 20: ...maximum output 5 About the display 1 Display after the decimal point Mode 00 P 19 Display 1 position of decimal point Display 2 setting of decimal point position 2 Clear the display Mode 08 P 31 33 Setting of blank display 3 Change the arithmetic measurement display value after reset setting of display offset P 44 4 Clear the last measurement data while the power is on Mode 09 P 34 35 Reset mode w...
Страница 21: ... fixed when pressing ENT key on the measurement display Measurement calculation method Decimal point position Setting the display digits which are after the decimal point Add the signal which was entered to A input when B input was in LOW level and deduct the signal which was entered to A input when B input was in HI level A B A B A or A Add the signal which was entered to A input and deduct the s...
Страница 22: ...0 0 Measurement calculation method 0 A B individual add subtract input Count value P20 ...
Страница 23: ...nt will be UP adding and Down subtraction will be reveresal Caution When this calculation method is chosen do not set alarm display selection and display 2 of analog output display selection If display 2 is set it might output irrespective of display Count value Count value Count value ON Up OFF Down P21 ...
Страница 24: ... data converter for integrating measurement By setting the 4digit value and EXP value which is set by Mode 02 scaling factor per 1 signal can be set until 1x10 9 9999 example Want to display integrating value as L using the flow sensor which is 2 5ml per 1 pulse Adjust the prefered unit L 4digits value EXP value Mode 01 Mode 02 P22 ...
Страница 25: ...e will not be erased example When the frequency divider is set to 003 1 3 outcome will be as follows 3 pulse output per 1 rotation and 0 5m roller sending per 1 rotation Using only scaling data converter might cause some error In that case divide the input Setting should be as follows Mode 01 0 1 5 0 0 0 0 5 5000x10 4 Mode 02 0 2 4 0 0 3 Frequency divider should be 3 as 1rotation produces 3pulse o...
Страница 26: ...ng procedure is same as Mode 01 A input Setting of scaling data converter Mode B Input Setting of EXP value and frequency divider 0 4 A B C D E F 0 4 3 0 0 1 frequency divider 3digits 1 1 1 999 000 to be 1 1000 EXP value 10 n n 0 9 Setting procedure is same as Mode 02 A input Setting of EXPvalue and frequency divider P24 ...
Страница 27: ...ing without re output Display selection 0 Display 1 1 Display 2 Caution Do not select display 2 when selecting A or A with Mode 00 measurement calculation method Display selection Display 1 Output against Display 1 Display 2 Output against Display 2 Preset action selection setting of output action For the timing chart and setting condition refer to output timing for preset action and setting condi...
Страница 28: ... PRESET B 0 PRESET B PRESET A OFFSET 1 Each 1stage preset action Setting name contents PRESET A Plus side setting value PRESET B Minus side setting value OFFSET Display value when reset display offset value Setting condition set each item as follow UP DOWN OFFSET 0 PRESET B Offset PRESET A Up Action Down Action Without automatic reset With automatic reset Without automatic reset With automatic res...
Страница 29: ... condition set each item as follow UP offset 0 Offset Preset A Preset B 0 DOWN Preset3 0 Offset Preset A Preset B 0 3 Plus 1stage and secondary output action Up Action Down Action Setting item Contents Preset A Pulse setting value Preset B Retardation setting value Offset Reset display value display offset setting value Setting condition set each item as follow UP offset 0 Offset Preset A 99 9 Pre...
Страница 30: ...thout automatic reset display value will not go back to display offset value when the output conditions are met With automatic reset dusplay value will go back to display offset value when the output conditions are met Actions of each preset are as follows 1 Plus side 2stages preset action UP reset Preset B display value display value display offset value recount DOWN reset Preset B display value ...
Страница 31: ... if the output conditions are met Reset input or automatic reset can activate the output with re output Sequence output Output will be ON when output only if condition are met However re output will be invalid when plus side 2 stages preset action with automatic reset is chosen 1 shot output 1 shot of pulse with set duration will be output when output conditions are met Caution When plus side 2 st...
Страница 32: ...tput 0 10 ms 5 250 ms 1 20 ms 6 500 ms 2 50 ms 7 750 ms 3 100 ms 8 1 sec 4 200 ms 9 2 sec Caution Output 1 3 s LED flashing might not be visible when 10ms is chosen although Out 1 3 Output mode 0 sequence output 1 1shot output OUT1 3 or OUT 2 4 output mode Sequence Output will be activated when the conditions are met If it is beyound tolerance condition output will be OFF 1 shot 1 shot of pulse wi...
Страница 33: ... measured value will blink when blank display is set Alarm output LED and hold LED will functuon normally External input function Setting the function of terminal block 2 to 3 When it is ON hold LED will blink 0 Banned input Bann the sensor input while ON 1 Hold input Keep displaying the current value while ON blink when it functions measurement is proceeding internally overscale LED will be kept ...
Страница 34: ... measurement 1 st round 1st overscale display OverscaleLED flashes Display will be ZERO with zero surporess 2nd overscale display 3 st round Display will be ZERO without zero surporess 3rd overscale display Display will blink at 999999 measurement continues internally 1 Endless Start measurement Overscale display Overscale LED flashes Display will be ZERO without zero surpress Run over 2 st round ...
Страница 35: ...ppear while pressing Up key and displaying overscale but if it exceed 99times over LED will blink Numbers of over under digits Display while Up key is on normal display Display over Run over Run over Zero surpress delete 0 in the upper digits sample Displaying 100 Display Display Without Zero surpress Uppler 2digits display over Surpressing Zero P33 ...
Страница 36: ...t Display1 Display2 Action Mode of Reset key 0 Not Reset 1 Reset in a bit 2 Need pushing key for 1 seconds 3 Need pushing key for 2 seconds Reset Node in Power ON 0 Not Reset 1 Reset Reset Node in Power ON Select record measured value at last time or delete measured value at last time in Power ON 0 You begin measuring from last measured value 1 The last measured value is deleted and You begin meas...
Страница 37: ...Display The integrated value is returned to display offset value The alarms of current display are canceled Display1 The integrated value of display1 is returned to display offset value The alarms of display1 are canceled Display2 The integrated value of display2 is returned to display offset value The alarms of display2 are canceled Count Input Signal ON OFF 2 Seconds P35 ...
Страница 38: ...tion 0 Standard 0 10V 0 5V 1 5V 4 20mA 1 Reverse 10 0V 5 0V 5 1V 20 4mA Analog Output Display Selection 0 Display1 1 Display2 Select of the Analog Output Display Select analog output display1 or analog output display2 Select of the Reverse Output Analog output is reversed Select of the Output Digits Analog Output 0 Synchronizing with indicated value Analog output to indicated value 1 Synchronizing...
Страница 39: ... decimal point at that time example 1 Analog output is AV5 0 10V Setting of maximum output when display is _ _ 1 0 0 0 A B C D E F Mode 10 1 0 0 E 0 A B C D E F Mode 11 1 0 1 0 0 0 C F Display is 1000 when output is maximum Caution Analog output is a absolute value of display value Output is as shown in the figure below example 2 Output digits is 4 digits of left Display is 2000 when maximum analo...
Страница 40: ...play Select BCD output display1 or analog output display2 BCD Data Output Timing 0 TI signal is used Capture prohibition signal 1 Request signal is used The update of data is demanded BCD Output Display selection Setting of the Output display data TI signal Parity logic Positive Logic The collector of output transistor and emitter are conducting Negative Logic The collector of output transistor an...
Страница 41: ...all 1 PRESET A 2 PRESET B Select of the BCD Data Input Select BCD input for which preset Latch Signal Input Logic Data is not input when latch signal is input 0 Latch on short Data is not input when latch signal pin and GND are short 1 Latch on open Data is not input when latch signal pin and GND are open BCD Data Input Logic Setting of the Logic of BCD Data which is inputed 0 Hi Active Data is in...
Страница 42: ...until return to measurment display 2 Mode Protect does not function 3 Reset it before starting measurement when PRESET Value is changed Display Operation Operation Key Press Mode key more than 2 seconds OUT1 3 s LED light then PRESET of OUT1 3 is called PRESET Value is switched pushing Mode Key Also The indicator A displays Change to right the position of flashing digit One hit makes 1digit move r...
Страница 43: ... Reset it before starting measurement when Display Offset Value is changed Display Operation Operation Key Press Mode key and Up key more than 2seconds D1LED s LED light then Display Offset of D1 is called PRESET Value is switched pushing Mode Key Also The indicator A displays Change to right the position of flashing digit One hit makes 1digit move right hand side Change the value of flashing digi...
Страница 44: ...can not be called during the setting of Mode Preset Display offset 2 Press down key more than 2seconds 3 Current mode protect is displayed after 2 seconds Mode protect ON A B C D E F L O N Mode protect OFF A B C D E F L O F F 4 In addition if press down key more than 8seconds mode protect function is changed ON changes into OFF and OFF changes into ON 5 Stop press down key and return to measuremen...
Страница 45: ...fer to Setting menu on page15 3 Adjust Span volume Zero volume to become the following values Adjust Span volume first Voltage output AV3 type Voltage output AV4 type Voltage output AV5 type Voltage output AI type Display Voltage 0 1V 10 5V Turn Zero volume Turn Span volume Display Voltage 0 0V Turn Zero volume 10 5V Turn Span volume Display Voltage 0 0V Turn Zero volume 10 10V Turn Span volume Di...
Страница 46: ...ping Do not touch excluding the emergency If you adjust sensitivity we do not take the responsibility Sinusoidal input type AC0 8 80Vp p Sine curve input type AC0 05 20Vp p Methode of adjustment 1 Remove attachment parts 2 You can see Sensitivity setting volume at ellipse s hole on left side of product See and adjust it Sensitivity High Low P47 ...
Страница 47: ...als to turning off logic of TI signal can be changed Refer to P40 4 When this option is selected D sub37P male connector XM2A 3701 and Hood XM2S 3711 are attached by the addition BCD Output Pin Layout D Sub 37P fenmale connector at meter side BCD Output Circuit Diagram NPN open collector pulse output BCD Output Time Chart Minus Sign Decimal Point TI Signal Parity Output REQ Input GND Meter side Pi...
Страница 48: ...al pin 37pin and GND are short 1 Latch on open Data is not input when latch signal pin 37pin and GND are open 4 When this option is selected D sub37P male connector XM2A 3701 and Hood XM2S 3711 are attached by the addition BCD Input Pin Layout D Sub 37P fenmale connector at meter side Taking Data Latch Signal Input logic is Latch on short Minus Sign Spare Input Meter side Latch Input GND Spare Inp...
Страница 49: ... cut dimensions and interval mm Attachement parts Rubber packing Accessory Use the rubber packing of Accessory if it is used in IP66 Terminal screw M3 5 Terminal width 7mm Terminal board is installed at shipping 130 or more 70 or more mm P50 ...
Страница 50: ...20 TERMINAL BOARD LABEL DRAWING NPN Open collector pulse Sinusoidal Electric voltage pulse Sine curve 90 phase contrast input Line receiver input mm P51 ...
Страница 51: ...21 UNIT LABEL DRAWING mm P52 ...
Страница 52: ...ING SPECIFICATIONS DT 601CG main unit with terminal board cover Air Cap packing material Air Cap packing material Poly Bag Rubber packing ACCESSORY Unit label D SUB37PIN male connector Hood B BI OPTION Box P53 ...
Страница 53: ... 2 Use three core shield line for sensor cord 3 Shorten sensor cord as much as possible 4 It is better not to connect machinery s ground wire with GND of meter 5 Use noise filter if there is a noise from power line Refer to following figure 6 wired method of sensor cord When the power cord passes near sensor cord sensor cord must be wired with single 7 When the noise is generated around product Us...
Страница 54: ... The display doesn t change Is setting of each mode correct Setting value is less than from 0 effective viewing area Is the sensor input normal Tighten the terminal screw again And check by test mode Refer to Page 15 Is the detection distance Confirm flashing sensor light of proximity sensor normal Are output signal of sensor and Confirm the specification sheet input method of meter correct Refer ...