6.0 Maintenance / Calibration
6.1
Calibration. On the main screen, there is a button in the upper right hand corner
labeled Calibration. Official Calibration can only be performed by an authorized SDL
Atlas technician and the area is password protected. Please contact your local agent
to arrange calibrations.
6.2
Care of Your Vortex M6
Use only a damp or sudsy cloth for cleaning the control panel. Some cleaning
products may harm the finish on the control panel or damage the interior. DO NOT
use products that contain alcohol on the control panel. Wipe the panel dry after
cleaning.
Wipe the instrument cabinet as needed. If detergent, bleach or other washing
products are spilled on the cabinet, wipe immediately. Some products will cause
permanent damage if spilled on the cabinet. Do not use scouring pads or abrasive
cleansers.
The washtub will need no particular care though it may need rinsing or wiping after
some unusual loads have been tested. This also may be necessary if too little
detergent has been used. Leave the lid open to allow the inside of the instrument to
dry out after use. This helps prevent musty odors from developing.
The agitator should not be removed except for service. The washtub is designed to
be self‐cleaning.
Cold Weather Care
If the instrument is delivered on a cold day (below freezing), or is stored in an
unheated room or area during the cold months, do not attempt to operate the
Vortex M6 until it has had a chance to warm up. Water from the previous cycle may
remain.
6.3
Replacing Hoses
Hoses and other natural rubber parts deteriorate after extended use. Hoses may
develop cracks, blisters or material wear from the temperature and constant high
pressure they are subjected to.
All hoses should be checked on a monthly basis for any visible signs of deterioration.
Any hose showing the signs of deterioration listed above should be replaced
immediately. All hoses should be replaced every five years.